Automatic Detection of Extracochlear Electrodes in Cochlear Implants Using Electric Field Imaging: Bench and Clinical Validation
Audiology Research · 14.08.2026 · 16 (4) · 119
DOI: 10.3390/audiolres16040119 ↗Kurz gefasst
Fachartikel in der Zeitschrift Audiology Research, online erschienen am 14.08.2026. Verfasst von Mehrangiz Ashiri und 18 weiteren Autor:innen. Beteiligte Einrichtung: Research and Technology, Advanced Bionics LLC, Valencia, CA 91355, USA. Die Arbeit nennt 28 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objectives
Extracochlear electrodes (EEs), defined as electrode contacts located outside the cochlea, can degrade cochlear implant (CI) performance. We present and validate an Electric Field Imaging (EFI)-based algorithm that helps detect EEs and can be used intra- or post-operatively.
Methods
An algorithm for detecting EEs from EFI data was developed. Validation and testing were performed using three datasets: (a) bench models with known EE conditions (saline and resistive load model), (b) clinical EFI recordings from CI recipients with imaging confirmation (CT or plain X-ray), and (c) EFI recordings without imaging. Primary outcomes included the ability to differentiate extracochlear conditions from fully inserted electrode arrays, as well as the concordance between the number of EE contacts identified by the algorithm and those determined from controlled bench configurations or available imaging data.
Results
The algorithm reliably differentiated full insertion from EE conditions on bench models and clinical EFI recordings from CI recipients with imaging confirmation. In the imaging-confirmed clinical cohort (6 EE cases among 226 CI recipients), the algorithm achieved 100% sensitivity and specificity. In clinical cases without imaging, the algorithm flagged 3.94% as having extracochlear electrodes, within the range of prevalence reported in the literature. Furthermore, the lateral-wall electrode arrays were associated with a significantly higher extracochlear electrode occurrence compared to the pre-curved arrays.
Conclusions
This EFI-based algorithm may provide a useful screening tool for EEs using routine clinical measurements, supporting intra-operative and post-operative detection while reducing reliance on imaging and thereby aiding clinical decision-making. The method can be integrated into existing clinical fitting software workflow and complements recent work on EFI-based tip fold-over detection. The observed sensitivity and specificity suggest that this approach may provide a valuable tool for identifying EEs in situations where imaging is unavailable or impractical.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Research and Technology, Advanced Bionics LLC, Valencia, CA 91355, USA.
- Auditory Neuroscience Lab, Barrow Neurological Institute, Phoenix, AZ 85013, USA.
- Department of Clinical Neurosciences, University of Cambridge, Cambridge CB2 0QQ, UK.
- Department of Otolaryngology, Washington University, St. Louis, MO 63130, USA.
- Department of Otolaryngology-Head and Neck Surgery, University of Utah, Salt Lake City, UT 84132, USA.
- Hearts for Hearing, Oklahoma City, OK 73120, USA.
ORCID-Profile: Mehrangiz Ashiri, Ashish Mehta, Daniele Borsetto, Susan T Eitutis, Jordan J Varghese
Zitieren
APA
Ashiri, M., Spahr, T., Chen, C., Botash, A., Mehta, A., Boyle, P., Bance, M., Borsetto, D., Eitutis, S. T., Varghese, J. J., Buchman, C. A., Gifford, R. H., DeFreese, A. J., Miller, M., Ahsan, S. F., Danner, C., Allen, K. P., Bartels, L., & Koka, K. (2026). Automatic Detection of Extracochlear Electrodes in Cochlear Implants Using Electric Field Imaging: Bench and Clinical Validation. Audiology Research, 16(4), 119. https://doi.org/10.3390/audiolres16040119
Vancouver
Ashiri M, Spahr T, Chen C, Botash A, Mehta A, Boyle P, et al. Automatic Detection of Extracochlear Electrodes in Cochlear Implants Using Electric Field Imaging: Bench and Clinical Validation. Audiology Research. 2026;16(4):119. doi:10.3390/audiolres16040119
BibTeX
@article{Ashiri2026,
author = {Ashiri, Mehrangiz and Spahr, Tony and Chen, Chen and Botash, Azret and Mehta, Ashish and Boyle, Patrick and Bance, Manohar and Borsetto, Daniele and Eitutis, Susan T and Varghese, Jordan J and Buchman, Craig A and Gifford, René H and DeFreese, Andrea J and Miller, Matthew and Ahsan, Syed F and Danner, Christopher and Allen, Kyle P and Bartels, Loren and Koka, Kanthaiah},
title = {Automatic Detection of Extracochlear Electrodes in Cochlear Implants Using Electric Field Imaging: Bench and Clinical Validation},
journal = {Audiology Research},
year = {2026},
volume = {16},
number = {4},
pages = {119},
doi = {10.3390/audiolres16040119}
}
EUVIMED