Measurement Performance of Electronic Radon Monitors

Thomas R. Beck, Elisabeth Foerster, Martin Biel, Sebastian Feige

Atmosphere · 30.09.2024 · 15 (10) · 1180

DOI: 10.3390/atmos15101180 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Atmosphere, online erschienen am 30.09.2024. Verfasst von Thomas R. Beck und 3 weiteren Autor:innen. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Euratom research and training programme 2014.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

The measurement performance and characteristics of electronic radon monitors with respect to radiological and environmental parameters are investigated. The study includes a sample of 14 different types of devices from nine manufacturers. The devices are currently available on the market with acquisition costs in the low or medium range. For comparison purposes, a high-end AlphaGUARD device is included in the study as a benchmark for measurement performance of radon monitors. Significant differences in the measurement performance are found between the tested instrument types. Overall, however, it can be concluded that most radon monitors perform acceptably and provide reliable information on radon activity concentrations in homes or workplaces, allowing residents and employers to make decisions about the need for radon protection measures. But it turns out that many radon monitors are supplied by the manufacturer with inadequate calibration, so that the instruments must be additionally calibrated in a reference atmosphere before they can be used. Among the tested radon monitors, there are also types with sufficiently good measuring performance, which represent an inexpensive alternative to high-end devices for radon professionals.

Förderung

  • Euratom research and training programme 2014

Zitieren

APA

Beck, T. R., Foerster, E., Biel, M., & Feige, S. (2024). Measurement Performance of Electronic Radon Monitors. Atmosphere, 15(10), 1180. https://doi.org/10.3390/atmos15101180

Vancouver

Beck TR, Foerster E, Biel M, Feige S. Measurement Performance of Electronic Radon Monitors. Atmosphere. 2024;15(10):1180. doi:10.3390/atmos15101180

BibTeX
@article{Beck2024,
  author = {Beck, Thomas R. and Foerster, Elisabeth and Biel, Martin and Feige, Sebastian},
  title = {Measurement Performance of Electronic Radon Monitors},
  journal = {Atmosphere},
  year = {2024},
  volume = {15},
  number = {10},
  pages = {1180},
  doi = {10.3390/atmos15101180}
}