EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Accuracy, Time, and Comfort of Different Intraoral Scanners: An In Vivo Comparison Study

Roberta Lione, Francesca Chiara De Razza, Francesca Gazzani, Letizia Lugli, Paola Cozza, Chiara Pavoni

Applied Sciences · 02.09.2024 · 14 (17) · 7731

DOI: 10.3390/app14177731 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Applied Sciences, online erschienen am 02.09.2024. Verfasst von Roberta Lione und 5 weiteren Autor:innen. Die Arbeit wurde bisher 13-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The purpose of this study was to compare the accuracy of three different intraoral scanners (IOSs) and to evaluate the patients’ experience.

Methods

Thirty subjects were scanned with three different IOSs (TRIOS Color®, iTero Element 5D®, and iTero Lumina®): a sample of 90 maxillary casts was collected. The inclusion criteria were permanent dentition, absence of defects during the dental impression, and complete record of the palatal vault. After the scanning procedure, patients were asked to answer a questionnaire (10 answers, scale from 1 to 10) to compare their experiences with iTero Element and iTero Lumina in the scanning procedure.

Results

Differences were recorded in the upper central incisor region (f-ratio of 4.186 for Mesiodistal 1.1 and f-ratio of 4.222 for Mesiodistal 2.1, p < 0.05), while intercanine width Upper 13–23 showed the smallest f-ratio (0.226) when compared to the other results, followed by Mesiodistal 1.4 (0.433). Patients reported more comfort, less pain, decreased duration, and a better technology visualization of the impression with iTero Lumina. No significant differences concerning dryness of the mouth and gag reflex were found.

Conclusions

The examined IOSs offer comparable accuracy in capturing dental arch dimensions. The interviewed patients expressed an overall preference for digital impressions performed with iTero Lumina, linked to increased comfort, painless practice, and a better technology visualization of the impression.

Zitieren

APA

Lione, R., Razza, F. C. D., Gazzani, F., Lugli, L., Cozza, P., & Pavoni, C. (2024). Accuracy, Time, and Comfort of Different Intraoral Scanners: An In Vivo Comparison Study. Applied Sciences, 14(17), 7731. https://doi.org/10.3390/app14177731

Vancouver

Lione R, Razza FCD, Gazzani F, Lugli L, Cozza P, Pavoni C. Accuracy, Time, and Comfort of Different Intraoral Scanners: An In Vivo Comparison Study. Applied Sciences. 2024;14(17):7731. doi:10.3390/app14177731

BibTeX
@article{Lione2024,
  author = {Lione, Roberta and Razza, Francesca Chiara De and Gazzani, Francesca and Lugli, Letizia and Cozza, Paola and Pavoni, Chiara},
  title = {Accuracy, Time, and Comfort of Different Intraoral Scanners: An In Vivo Comparison Study},
  journal = {Applied Sciences},
  year = {2024},
  volume = {14},
  number = {17},
  pages = {7731},
  doi = {10.3390/app14177731}
}