Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation

Tairan Ji, Teng Jing, Jianan Cheng

Frontiers in Physiology · 18.12.2025 · 16

DOI: 10.3389/fphys.2025.1632333 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Physiology, online erschienen am 18.12.2025. Verfasst von Tairan Ji und 2 weiteren Autor:innen. Beteiligte Einrichtung: National Research Center of Pumps, Jiangsu University, Zhenjiang, Jiangsu, China. Die Arbeit wurde bisher 2-mal zitiert (Crossref) und nennt 32 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Traditional percutaneous ventricular assist devices (PVADs) face limitations due to their small implantation size, requiring higher rotational speeds to meet left ventricular support demands. However, the elevated shear stress induced by high-speed operation leads to excessive hemolysis, necessitating design improvements for better clinical applicability.

Method

This study presents a foldable implantable blood pump featuring a collapsible flexible impeller and pump casing, which reduces its profile during implantation and expands to operational size at the target position, thereby allowing lower rotational speeds and improved hemolytic performance. Through numerical simulations combined with response surface methodology (RSM), we systematically analyzed the influence of various structural parameters on pump head and hemolysis index (HI), aiming at parametric optimization of the initial model for enhanced hydraulic efficiency and reduced blood trauma risk.

Results

Simulation analysis of seven parameter configurations identified five key structural parameters with dominant effects on head and HI: impeller inlet angle, impeller outlet angle, diffuser inlet angle, diffuser wrap angle, and impeller-diffuser gap. Using Box-Behnken Design (BBD), we established dual-response prediction models for both head and HI, analyzed significant interaction terms, and derived an optimized configuration achieving a 6.9% increase in head pressure and 17.9% reduction in HI compared to the initial design. Additional analysis revealed that increasing tip clearance reduces hemolysis at the cost of moderate head decrease.

Conclusion

We designed and optimized a foldable axial-flow blood pump for transaortic implantation. The optimized configuration demonstrates a 6.9% head improvement to 2.346 m and 17.9% HI reduction to 1.081 × 1 0 − 2 % . Furthermore, increasing tip clearance from 0.2 mm to 0.4 mm provides additional hemolysis reduction without excessively compromising the pressure head.

Schlagwörter

Stichwörter der Autor:innen

LVADcage filamentcomputational hemodynamicsfoldable blood pumpparametricoptimizationresponse surface methodology

Autor:innen und Institutionen

  • National Research Center of Pumps, Jiangsu University, Zhenjiang, Jiangsu, China.

Zitieren

APA

Ji, T., Jing, T., & Cheng, J. (2025). Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation. Frontiers in Physiology, 16. https://doi.org/10.3389/fphys.2025.1632333

Vancouver

Ji T, Jing T, Cheng J. Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation. Frontiers in Physiology. 2025;16. doi:10.3389/fphys.2025.1632333

BibTeX
@article{Ji2025,
  author = {Ji, Tairan and Jing, Teng and Cheng, Jianan},
  title = {Mechanically-foldable axial flow blood pump: response-surface-based structural optimization and hemolytic performance evaluation},
  journal = {Frontiers in Physiology},
  year = {2025},
  volume = {16},
  doi = {10.3389/fphys.2025.1632333}
}
Kostenlos starten →