Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system

Yuting Lin, Wangyao Li, Aoxiang Wang, Daniel Johnson, Gregory N Gan, Hao Gao

Frontiers in Oncology · 19.07.2024 · 14

DOI: 10.3389/fonc.2024.1421869 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Oncology, online erschienen am 19.07.2024. Verfasst von Yuting Lin und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, KS, United States. Die Arbeit wurde bisher 25-mal zitiert (Crossref) und nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: NCI NIH HHS, National Institute for Health and Care Research, Cancer Center, University of Kansas.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Proton minibeam radiation therapy (pMBRT) can deliver spatially fractionated dose distributions with submillimeter resolution. These dose distributions exhibit significant heterogeneity in both depth and lateral directions. Accurate characterization of pMBRT doses requires dosimetry devices with high spatial resolution and a wide dynamic range. Furthermore, the dependency of dosimetric measurements on Linear Energy Transfer (LET), as observed in conventional proton therapy, is also present in pMBRT depth dose measurements.

Purpose

This work demonstrates the process of performing comprehensive dosimetric measurements to characterize the pMBRT collimator on a clinical single-gantry proton machine, utilizing commercially available dosimetry devices.

Methods

The minibeam collimator is designed to be mounted on the clinical nozzle as a beam-modifying accessory. Three collimators, each with a slit opening of 0.4 mm, are thoroughly evaluated. The center-to-center (c-t-c) distances of the slits for these collimators are 2.8 mm, 3.2 mm, and 4.0 mm, respectively. High spatial resolution dosimetry devices are essential for PMBRT dose characterizations. To meet this requirement, two-dimensional (2D) dose measurement devices, Gafchromic films, are used to measure lateral profiles at various depths. Films are also used for depth dose profile measurements in solid water. Additionally, high-resolution point dose detectors, microDiamond, and Razor diode detectors are employed for lateral profile measurements at various depths. Percent depth dose (PDD) measurements of pMBRT in solid water, with various proton energies, collimators, and air gaps, are performed using Gafchromic films. The film’s LET dependency for proton beams is corrected to ensure accurate pMBRT PDD measurements. The Monte Carlo simulation tool TOPAS is utilized to compare and validate all experimental measurements.

Results

At depths where LET is not a concern, film dose measurements were consistent with microDiamond and Razor diode point measurements. The point detectors need to be orientated with the thin side aligned to the incoming beam. Comparison of the lateral dose profiles extracted from TOPAS simulations, films, microDiamond, and Razor diode detectors shows a passing rate exceeding 98% in 1D gamma analysis at 3% 0.1 mm criteria.However, when the microDiamond detector is orientated to face the pMBRT beam, its spatial resolution may not be sufficient to capture the peak and valley dose accurately. Nevertheless, an accuracy within 2% can still be achieved when comparing the average dose. The PDD measurements show that the peak valley dose ratio (PVDR) of pMBRT can be altered at different depths with different air gaps using the same collimator or different collimators of different c-t-c distances.

Conclusion

Our study demonstrates that comprehensive dose measurements for pMBRT can be conducted using standard clinical dose measurement devices. These measurements are indispensable for guiding and ensuring accurate dose reporting in pre-clinical studies using the pMBRT technique.

Schlagwörter

Stichwörter der Autor:innen

film dosimetryproton dosimetryproton mini-beamproton therapysmall field dosimetryspatially fractionated radiotherapy

Autor:innen und Institutionen

  • Department of Radiation Oncology, University of Kansas Medical Center, Kansas City, KS, United States.
  • Department of Biomedical Engineering, Huazhong University of Science and Technology, Wuhan, China.

Förderung

  • NCI NIH HHS
  • National Institute for Health and Care Research
  • Cancer Center, University of Kansas

Zitieren

APA

Lin, Y., Li, W., Wang, A., Johnson, D., Gan, G. N., & Gao, H. (2024). Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system. Frontiers in Oncology, 14. https://doi.org/10.3389/fonc.2024.1421869

Vancouver

Lin Y, Li W, Wang A, Johnson D, Gan GN, Gao H. Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system. Frontiers in Oncology. 2024;14. doi:10.3389/fonc.2024.1421869

BibTeX
@article{Lin2024,
  author = {Lin, Yuting and Li, Wangyao and Wang, Aoxiang and Johnson, Daniel and Gan, Gregory N and Gao, Hao},
  title = {Comprehensive dosimetric commissioning of proton minibeam radiotherapy on a single gantry proton system},
  journal = {Frontiers in Oncology},
  year = {2024},
  volume = {14},
  doi = {10.3389/fonc.2024.1421869}
}
Kostenlos starten →