Assessment of bioelectrical impedance analysis devices for data reliability of body impedance measurements
Frontiers in Nutrition · 13.01.2026 · 12
DOI: 10.3389/fnut.2025.1705346 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Nutrition, online erschienen am 13.01.2026. Verfasst von Dan Bee Kim und 2 weiteren Autor:innen. Beteiligte Einrichtung: Korea Research Institute of Standards and Science, Daejeon, Republic of Korea. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 42 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Bioelectrical impedance analysis (BIA) devices are widely used for body composition analyses, but their accuracy for body impedance measurements can vary significantly between BIA device models. Calibration of BIA device is essential to ensure the reliability of measurement data, particularly for clinical and research purposes. This study aims to evaluate the electric impedance measurement abilities of multi-frequency BIA devices using a home-built body impedance imitator (BII), designed to simulate human body impedance, and to assess the measurement uncertainty based on calibration with internationally accredited reference standards.
Methods
Two different models of multi-frequency BIA device with similar specifications and measurement schemes were evaluated using the BII as a reference standard. First, the impedance values of the BII were calibrated using an LCR meter, traceable to national impedance standards. Next, the impedance measurements were performed using both models based on the standard BII over a wide frequency range (1 kHz to 1 MHz) to analyze the accuracy, linearity, contact resistance effects, etc. of the BIA devices.
Results
Despite having similar specifications, two models of BIA device exhibited notable differences in their measurement performance. Model A showed better accuracy and consistency, with measurement uncertainty of approximately 1.4% at a 95% confidence level ( k = 2). In contrast, Model B showed higher deviations, particularly in trunk impedance measurements. The study found that the BIA device calibration procedure could effectively quantify the measurement uncertainty and ensure the reliability of measurement data provided by the BIA device.
Conclusion
This study demonstrates that BIA devices can be calibrated using a standard procedure with a stable reference such as the BII, which traces back to internationally accredited impedance standards. The results show that the calibration significantly improves the reliability of BIA device measurements, providing a level of confidence for the measurement data.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Korea Research Institute of Standards and Science, Daejeon, Republic of Korea.
Zitieren
APA
Kim, D. B., Shin, S. S., & Kim, W. S. (2026). Assessment of bioelectrical impedance analysis devices for data reliability of body impedance measurements. Frontiers in Nutrition, 12. https://doi.org/10.3389/fnut.2025.1705346
Vancouver
Kim DB, Shin SS, Kim WS. Assessment of bioelectrical impedance analysis devices for data reliability of body impedance measurements. Frontiers in Nutrition. 2026;12. doi:10.3389/fnut.2025.1705346
BibTeX
@article{Kim2026,
author = {Kim, Dan Bee and Shin, Seong Su and Kim, Wan-Seop},
title = {Assessment of bioelectrical impedance analysis devices for data reliability of body impedance measurements},
journal = {Frontiers in Nutrition},
year = {2026},
volume = {12},
doi = {10.3389/fnut.2025.1705346}
}
EUVIMED