A exploratory pilot method comparison study of two viscoelastic coagulation analyzers
Frontiers in Medicine · 07.08.2026 · 13
DOI: 10.3389/fmed.2026.1874412 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Medicine, online erschienen am 07.08.2026. Verfasst von Daniel Ricardo Argumedo und 2 weiteren Autor:innen. Beteiligte Einrichtung: Center of Natural and Human Science, Federal University of ABC, Santo André, Brazil. Die Arbeit nennt 21 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
This preliminary exploratory study evaluated the analytical performance of two different brand of viscoelastic coagulation analyzers in real-time blood samples from ten anonymous patients were analyzed on both devices, and the results were compared using Spearman correlation, Wilcoxon paired testing, Passing–Bablok regression, Bland–Altman agreement analysis, and Lin's concordance correlation coefficient. Systematic biases and variable concordance were observed across several parameters, indicating that the current exploratory dataset does not support direct numerical interchangeability between the two platforms without further testing in larger groups. As new platforms enter the global market, method-comparison studies using agreement-based statistical frameworks should be considered.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Center of Natural and Human Science, Federal University of ABC, Santo André, Brazil.
- Engineering, Modelling and Applied Social Sciences Center, Federal University of ABC, São Bernardo do Campo, Brazil.
- Department of Electrical Engineering, Universidad Autónoma Metropolitana-Iztapalapa, Mexico City, Mexico.
Zitieren
APA
Ricardo Argumedo, D., Maranho, J. G., & Ortiz-Posadas, M. R. (2026). A exploratory pilot method comparison study of two viscoelastic coagulation analyzers. Frontiers in Medicine, 13. https://doi.org/10.3389/fmed.2026.1874412
Vancouver
Ricardo Argumedo D, Maranho JG, Ortiz-Posadas MR. A exploratory pilot method comparison study of two viscoelastic coagulation analyzers. Frontiers in Medicine. 2026;13. doi:10.3389/fmed.2026.1874412
BibTeX
@article{RicardoArgumedo2026,
author = {Ricardo Argumedo, Daniel and Maranho, Júlio Gallinaro and Ortiz-Posadas, Martha R},
title = {A exploratory pilot method comparison study of two viscoelastic coagulation analyzers},
journal = {Frontiers in Medicine},
year = {2026},
volume = {13},
doi = {10.3389/fmed.2026.1874412}
}
EUVIMED