Phantom-based evaluation of radiomics feature stability for low-dose CT lung cancer screening
Frontiers in Endocrinology · 21.07.2026 · 17
DOI: 10.3389/fendo.2026.1880461 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Endocrinology, online erschienen am 21.07.2026. Verfasst von Sunyi Zheng und 15 weiteren Autor:innen. Beteiligte Einrichtung: Department of Radiology, Medical Artificial General Intelligence for Computation (MAGIC) Lab, National Clinical Research Center for Cancer, Tianjin's…. Hauptthemen laut MEDLINE: Radiomics, Lung Neoplasms, Tomography, X-Ray Computed, Phantoms, Imaging. Die Arbeit nennt 39 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
To assess the stability of radiomic features derived from lung nodules under low-dose CT lung cancer screening conditions and to evaluate the influence of feature stability on model performance.
Methods
A chest phantom containing eight simulated lung nodules was scanned on five CT scanners using varying tube voltages (100–120 kVp) and tube currents (20–60 mA·s) to simulate inter-scanner and intra-scanner variability. Nodule radiomic features were extracted accordingly and their stability was evaluated using the intraclass correlation coefficient. Stable features were grouped by hierarchical clustering, from which representative stable features were selected for each cluster. Models constructed using representative stable features and remaining unstable features were compared in two independent lung cancer screening datasets for nodule malignancy assessment and growth prediction. Model performance was evaluated using the area under the receiver operating characteristic curve (AUC).
Results
Inter-scanner variability had a greater impact on feature stability than intra-scanner variability, with tube current exerting more influence than tube voltage. Models based on representative stable features achieved better performance than those using unstable features in both the malignancy assessment cohort (AUC, 0.995 vs 0.945) and nodule growth prediction cohort (AUC, 0.799 vs 0.610). Models using representative stable features also showed smaller performance differences between training and test sets than those using unstable features.
Conclusion
Radiomic features with high stability under lung cancer screening conditions were associated with improved performance consistency in independent screening datasets. These findings suggest stability-informed feature selection may contribute to more reliable radiomics applications in lung cancer screening.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Radiology, Medical Artificial General Intelligence for Computation (MAGIC) Lab, National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, State Key Laboratory of Druggabi…
- Department of Radiology, Huadu District People's Hospital of Guangzhou, Guangzhou, Guangdong, China.
- Academy of Medical Engineering and Translational Medicine, Tianjin University, Tianjin, China.
- National Clinical Research Center for Cancer, Tianjin's Clinical Research Center for Cancer, State Key Laboratory of Druggability Evaluation and Systematic Translational Medicine, Tianjin Key Laboratory of Digestive Can…
- Department of Radiology, Ordos Central Hospital, Ordos, China.
- Department of Public Health, Erasmus Medical Center Rotterdam, Rotterdam, Netherlands.
Zitieren
APA
Zheng, S., Yang, X., Wang, H., Gao, Z., Ye, P., Wang, W., Li, W., Meng, D., Zhang, S., Zhang, W., Liu, H., Huang, S., Zhang, C., Wang, J., Hao, J., & Cui, X. (2026). Phantom-based evaluation of radiomics feature stability for low-dose CT lung cancer screening. Frontiers in Endocrinology, 17. https://doi.org/10.3389/fendo.2026.1880461
Vancouver
Zheng S, Yang X, Wang H, Gao Z, Ye P, Wang W, et al. Phantom-based evaluation of radiomics feature stability for low-dose CT lung cancer screening. Frontiers in Endocrinology. 2026;17. doi:10.3389/fendo.2026.1880461
BibTeX
@article{Zheng2026,
author = {Zheng, Sunyi and Yang, Xiaomeng and Wang, Hongren and Gao, Zhipeng and Ye, Pengchun and Wang, Weiping and Li, Wenhua and Meng, Donghua and Zhang, Shuai and Zhang, Wenjia and Liu, Houpu and Huang, Shuyuan and Zhang, Chunlin and Wang, Jing and Hao, Jihui and Cui, Xiaonan},
title = {Phantom-based evaluation of radiomics feature stability for low-dose CT lung cancer screening},
journal = {Frontiers in Endocrinology},
year = {2026},
volume = {17},
doi = {10.3389/fendo.2026.1880461}
}
EUVIMED