Thickness distribution and site-specific variability of thermoformed multilayer clear aligners: an in vitro study
Frontiers in Dental Medicine · 01.06.2026 · 7
DOI: 10.3389/fdmed.2026.1863922 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Dental Medicine, online erschienen am 01.06.2026. Verfasst von Thi Bich Van Tran und 1 weiteren Autor:innen. Beteiligte Einrichtung: Faculty of Dentistry, University of Medicine and Pharmacy, Ho Chi Minh City, Vietnam. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 17 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Thermoforming is a critical step in the fabrication of clear aligners and is associated with non-uniform deformation of thermoplastic materials, resulting in non-uniform thickness distribution. Because aligner thickness directly influences stiffness and force delivery, such variability may affect the accuracy and predictability of orthodontic treatment.
Aim
This study aimed to evaluate the thickness distribution of thermoformed clear aligners across different tooth regions and anatomical sites, with a focus on identifying patterns of non-uniform material thinning.
Methods
Ten thermoformed aligners fabricated from 760 µm multilayer thermoplastic sheets (Zendura FLX) were analyzed in vitro . Thickness measurements were obtained using a calibrated digital caliper at standardized anatomical reference points across anterior, canine, premolar, and molar regions. Generalized estimating equations (GEE) with robust standard errors were used to account for repeated measurements within the same aligner, and pairwise comparisons were adjusted using the Bonferroni correction.
Results
Significant differences in thickness were observed among tooth regions ( p < 0.001), with the lowest values in anterior teeth (404.0 ± 5.9 µm) and the highest in molars (481.6 ± 6.0 µm). Thickness reduction ranged from approximately 36% to 47% relative to the original material. Site-specific analysis demonstrated that the facial regions generally exhibited lower thickness values than corresponding palatal regions, indicating non-uniform thermoforming-induced deformation patterns. These findings demonstrated that anatomical location was a major determinant of thickness variability.
Conclusion
Thermoformed clear aligners exhibit pronounced and systematic non-uniform thickness distribution strongly influenced by tooth morphology and local geometry characteristics. This non-uniform thickness distribution may contribute to variations in force delivery and should be considered in aligner design and clinical application.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Faculty of Dentistry, University of Medicine and Pharmacy, Ho Chi Minh City, Vietnam.
Zitieren
APA
Tran, T. B. V., & Pham, N. A. T. (2026). Thickness distribution and site-specific variability of thermoformed multilayer clear aligners: an in vitro study. Frontiers in Dental Medicine, 7. https://doi.org/10.3389/fdmed.2026.1863922
Vancouver
Tran TBV, Pham NAT. Thickness distribution and site-specific variability of thermoformed multilayer clear aligners: an in vitro study. Frontiers in Dental Medicine. 2026;7. doi:10.3389/fdmed.2026.1863922
BibTeX
@article{Tran2026,
author = {Tran, Thi Bich Van and Pham, Ngoc Anh Thu},
title = {Thickness distribution and site-specific variability of thermoformed multilayer clear aligners: an in vitro study},
journal = {Frontiers in Dental Medicine},
year = {2026},
volume = {7},
doi = {10.3389/fdmed.2026.1863922}
}
EUVIMED