EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Intra-individual comparison of coronary calcium scoring between photon counting detector- and energy integrating detector-CT: Effects on risk reclassification

Elias V Wolf, Moritz C Halfmann, U Joseph Schoepf, Emese Zsarnoczay, Nicola Fink, Joseph P Griffith, Gilberto J Aquino, Martin J Willemink, Jim O'Doherty, Michaela M Hell, Pal Suranyi, Ismael M Kabakus und 3 weitere

Frontiers in Cardiovascular Medicine · 19.01.2023 · 9

DOI: 10.3389/fcvm.2022.1053398 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Cardiovascular Medicine, online erschienen am 19.01.2023. Verfasst von Elias V Wolf und 14 weiteren Autor:innen. Beteiligte Einrichtung: Department of Diagnostic and Interventional Radiology, University Medical Center of the Johannes Gutenberg-University, Mainz, Germany. Die Arbeit wurde bisher 28-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0). Gefördert von: Siemens Healthineers.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To compare coronary artery calcium volume and score (CACS) between photon-counting detector (PCD) and conventional energy integrating detector (EID) computed tomography (CT) in a phantom and prospective patient study.

Methods

A commercially available CACS phantom was scanned with a standard CACS protocol (120 kVp, slice thickness/increment 3/1.5 mm, and a quantitative Qr36 kernel), with filtered back projection on the EID-CT, and with monoenergetic reconstruction at 70 keV and quantum iterative reconstruction off on the PCD-CT. The same settings were used to prospectively acquire data in patients ( n = 23, 65 ± 12.1 years), who underwent PCD- and EID-CT scans with a median of 5.5 (3.0–12.5) days between the two scans in the period from August 2021 to March 2022. CACS was quantified using a commercially available software solution. A regression formula was obtained from the aforementioned comparison and applied to simulate risk reclassification in a pre-existing cohort of 514 patients who underwent a cardiac EID-CT between January and December 2021.

Results

Based on the phantom experiment, CACS PCD–CT showed a more accurate measurement of the reference CAC volumes (overestimation of physical volumes: PCD-CT 66.1 ± 1.6% vs. EID-CT: 77.2 ± 0.5%). CACS EID–CT and CACS PCD–CT were strongly correlated, however, the latter measured significantly lower values in the phantom (CACS PCD–CT : 60.5 (30.2–170.3) vs CACS EID–CT 74.7 (34.6–180.8), p = 0.0015, r = 0.99, mean bias –9.7, Limits of Agreement (LoA) –36.6/17.3) and in patients (non-significant) (CACS PCD–CT : 174.3 (11.1–872.7) vs CACS EID–CT 218.2 (18.5–876.4), p = 0.10, r = 0.94, mean bias –41.1, LoA –315.3/232.5). The systematic lower measurements of Agatston score on PCD-CT system led to reclassification of 5.25% of our simulated patient cohort to a lower classification class.

Conclusion

CACS PCD–CT is feasible and correlates strongly with CACS EID–CT , however, leads to lower CACS values. PCD-CT may provide results that are more accurate for CACS than EID-CT.

Schlagwörter

Stichwörter der Autor:innen

coronary artery calcium scoring (CACS)coronary artery diseaseenergy-integrating detector computed tomographyphoton-counting detector computed tomographyrisk stratification

Autor:innen und Institutionen

  • Department of Diagnostic and Interventional Radiology, University Medical Center of the Johannes Gutenberg-University, Mainz, Germany.
  • Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, Charleston, SC, United States.
  • German Centre for Cardiovascular Research, Partner Site Rhine-Main, Mainz, Germany.
  • MTA-SE Cardiovascular Imaging Research Group, Medical Imaging Center, Semmelweis University, Budapest, Hungary.
  • Department of Radiology, University Hospital Munich, LMU Munich, Munich, Germany.
  • Department of Radiology, Stanford University School of Medicine, Stanford, CA, United States.

Förderung

  • Siemens Healthineers

Zitieren

APA

Wolf, E. V., Halfmann, M. C., Schoepf, U. J., Zsarnoczay, E., Fink, N., Griffith, J. P., Aquino, G. J., Willemink, M. J., O'Doherty, J., Hell, M. M., Suranyi, P., Kabakus, I. M., Baruah, D., Varga-Szemes, A., & Emrich, T. (2023). Intra-individual comparison of coronary calcium scoring between photon counting detector- and energy integrating detector-CT: Effects on risk reclassification. Frontiers in Cardiovascular Medicine, 9. https://doi.org/10.3389/fcvm.2022.1053398

Vancouver

Wolf EV, Halfmann MC, Schoepf UJ, Zsarnoczay E, Fink N, Griffith JP, et al. Intra-individual comparison of coronary calcium scoring between photon counting detector- and energy integrating detector-CT: Effects on risk reclassification. Frontiers in Cardiovascular Medicine. 2023;9. doi:10.3389/fcvm.2022.1053398

BibTeX
@article{Wolf2023,
  author = {Wolf, Elias V and Halfmann, Moritz C and Schoepf, U Joseph and Zsarnoczay, Emese and Fink, Nicola and Griffith, Joseph P and Aquino, Gilberto J and Willemink, Martin J and O'Doherty, Jim and Hell, Michaela M and Suranyi, Pal and Kabakus, Ismael M and Baruah, Dhiraj and Varga-Szemes, Akos and Emrich, Tilman},
  title = {Intra-individual comparison of coronary calcium scoring between photon counting detector- and energy integrating detector-CT: Effects on risk reclassification},
  journal = {Frontiers in Cardiovascular Medicine},
  year = {2023},
  volume = {9},
  doi = {10.3389/fcvm.2022.1053398}
}
Kostenlos starten →