Analytical performance evaluation of MAGLUMI HBeAg and anti-HBe chemiluminescence immunoassays and comparison with the Abbott assays
Frontiers in Cellular and Infection Microbiology · 19.08.2026 · 16
DOI: 10.3389/fcimb.2026.1867239 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Frontiers in Cellular and Infection Microbiology, online erschienen am 19.08.2026. Verfasst von Jingfeng Zeng und 8 weiteren Autor:innen. Beteiligte Einrichtung: Blood Screening Department, Shenzhen Blood Center, Shenzhen, China. Hauptthemen laut MEDLINE: Immunoassay, Hepatitis B e Antigens, Hepatitis B. Die Arbeit nennt 35 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background
Hepatitis B e antigen (HBeAg) and its corresponding antibody (anti-HBe) serve as critical serological markers in the evaluation and therapeutic management of Hepatitis B virus (HBV) infection. Study design This retrospective study was designed to evaluate the performance of the newly CE-marked MAGLUMI HBeAg (CLIA) and MAGLUMI Anti-HBe (CLIA) assays by comparing with Abbott Architect HBeAg and Abbott Architect Anti-HBe assays, respectively. Analytical performance was verified in compliance with the guidelines issued by the Clinical and Laboratory Standards Institute (CLSI). Negative and positive specimens were collected to determine negative percent agreement (NPA) and positive percent agreement (PPA) respectively. Analytical specificity was assessed using negative samples containing potentially cross-reacting substances. Additionally, analytical sensitivity was evaluated using 1st WHO International Standard.
Results
The claimed performance specifications for LoB, LoD, linearity, and precision were successfully verified. Both MAGLUMI HBeAg (CLIA) and MAGLUMI Anti-HBe (CLIA) assays exhibited satisfactory NPA and PPA (≥98%). No significant cross-reactivity or interference was observed. Furthermore, all tested plasma matrices demonstrated equivalence to serum in both assays.
Conclusions
The MAGLUMI HBeAg (CLIA) and MAGLUMI Anti-HBe (CLIA) met the CLIS verification criteria with favorable analytical performance, and achieved high NPA and PPA relative to the Abbott Architect comparative assays. Besides, future concordance assessments based on a large intended cohort, together with independent clinical validation, are required to fully confirm their suitability for routine clinical application and management of HBV infection.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Blood Screening Department, Shenzhen Blood Center, Shenzhen, China.
- Medical Research and Laboratory Diagnostic Center, Jinan Central Hospital Affiliated to Shandong First Medical University, Jinan, China.
- Research & Development Department, Shenzhen New Industries Biomedical Engineering Co., Ltd. (Snibe), Shenzhen, China.
Zitieren
APA
Zeng, J., Zhao, X., Wang, W., Wang, J., Li, X., Zheng, K., Zhang, H., Yin, J., & Fang, Z. (2026). Analytical performance evaluation of MAGLUMI HBeAg and anti-HBe chemiluminescence immunoassays and comparison with the Abbott assays. Frontiers in Cellular and Infection Microbiology, 16. https://doi.org/10.3389/fcimb.2026.1867239
Vancouver
Zeng J, Zhao X, Wang W, Wang J, Li X, Zheng K, et al. Analytical performance evaluation of MAGLUMI HBeAg and anti-HBe chemiluminescence immunoassays and comparison with the Abbott assays. Frontiers in Cellular and Infection Microbiology. 2026;16. doi:10.3389/fcimb.2026.1867239
BibTeX
@article{Zeng2026,
author = {Zeng, Jingfeng and Zhao, Xin and Wang, Wenfeng and Wang, Ju and Li, Ximiao and Zheng, Kaize and Zhang, Hongwei and Yin, Jun and Fang, Zhonggang},
title = {Analytical performance evaluation of MAGLUMI HBeAg and anti-HBe chemiluminescence immunoassays and comparison with the Abbott assays},
journal = {Frontiers in Cellular and Infection Microbiology},
year = {2026},
volume = {16},
doi = {10.3389/fcimb.2026.1867239}
}
EUVIMED