High‐speed centrifugation rather than Lipoclear reagent can be used for removing the interference of lipemia on serological tests of infectious diseases: AIDS, hepatitis B, hepatitis C, and syphilis by chemiluminescent microparticle immunoassay
Journal of Medical Virology · 2024
DOI: 10.1002/jmv.29385 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Medical Virology, 2024. Verfasst von Weiping Liu und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Laboratory, Zigong First People's Hospital, Zigong, Sichuan, China. Hauptthemen laut MEDLINE: Syphilis, Acquired Immunodeficiency Syndrome, Hepatitis B, Hyperlipidemias. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 16 Literaturangaben. Gefördert von: Foundation of Key Science and Technology Project of Zigong of China, Clinical Research and Transformation Fund of Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, Health Care Committee of Sichuan Province.
Abstract (Originalsprache)
The aim of this study was to investigate the interference of lipemia on measurement of HBsAg, anti‐HBs, HBeAg, anti‐HBe, anti‐HBc, anti‐HCV, HIV Ag/Ab, and anti‐TP in serum by chemiluminescent microparticle immunoassay (CMIA) and compare lipemia removing performance between high‐speed centrifugation and Lipoclear reagent. Mixed native serum samples (NSs) and hyperlipemia serum samples (HLS) were prepared for the investigated parameters. The levels of these parameters in NS and HLS were determined by CMIA on an Abbott ARCHITECT i2000SR immunoassay analyzer. HBsAg, anti‐HBs, and anti‐TP were affected with relative bias >12.5% (acceptable limit) when the level of triacylglycerol (TG) was higher than 27.12 mmol/L in HLS. Clinically unacceptable bias were observed for HBeAg and anti‐HBe in HLS with TG higher than 40.52 mmol/L. However, anti‐HCV and HIV Ag/Ab were not interfered in severe lipemia with TG < 52.03 mmol/L. In addition, the Lipoclear reagent did not reduce the interference of lipemia with relative bias from −62.50% to −18.02%. The high‐speed centrifugation under the optimized condition of 12 000 g for 10 min successfully removed the interference of lipemia with relative bias from −5.93% to 0% for HBsAg, anti‐HBs, HBeAg, anti‐HBe, anti‐HBc, and anti‐TP. To conclude, high‐speed centrifugation can be used for removing the interference of lipemia to measure HBsAg, anti‐HBs, HBeAg, anti‐HBe, anti‐HBc, and anti‐TP. Accordingly, a standardized sample preanalytical preparation of the patients and other screening participants as well as a specimen examination procedure for removing lipemia interference on the serological tests was recommended.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Department of Clinical Laboratory, Zigong First People's Hospital, Zigong, Sichuan, China.
- Department of Laboratory Medicine, Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital, University of Electronic Science and Technology of China, Chengdu, Sichuan, China.
ORCID-Profile: Weiping Liu, Yaohui Song, Xia Long
Förderung
- Foundation of Key Science and Technology Project of Zigong of China
- Clinical Research and Transformation Fund of Sichuan Academy of Medical Sciences & Sichuan Provincial People's Hospital
- Health Care Committee of Sichuan Province
- Sichuan Science and Technology Program
Zitieren
APA
Liu, W., Li, L., Yin, M., Cao, C., Song, Y., & Long, X. (2024). High‐speed centrifugation rather than Lipoclear reagent can be used for removing the interference of lipemia on serological tests of infectious diseases: AIDS, hepatitis B, hepatitis C, and syphilis by chemiluminescent microparticle immunoassay. Journal of Medical Virology. https://doi.org/10.1002/jmv.29385
Vancouver
Liu W, Li L, Yin M, Cao C, Song Y, Long X. High‐speed centrifugation rather than Lipoclear reagent can be used for removing the interference of lipemia on serological tests of infectious diseases: AIDS, hepatitis B, hepatitis C, and syphilis by chemiluminescent microparticle immunoassay. Journal of Medical Virology. 2024. doi:10.1002/jmv.29385