Three-dimensional finite element analysis of translaminar lag screw designs for L5 bilateral spondylolysis
Frontiers in Bioengineering and Biotechnology · 13.08.2026 · 14
DOI: 10.3389/fbioe.2026.1882221 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 13.08.2026. Verfasst von Feng Li und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Spine Surgery, The 940th Hospital of Joint Logistic Support Force of Chinese People's Liberation Army, Lanzhou, Gansu, China. Die Arbeit nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
This finite element study aimed to quantitatively compare the biomechanical properties of translaminar lag screws with different diameters and structural designs for the surgical repair of L5 bilateral spondylolysis, to provide evidence for individualized screw selection.
Methods
An intact model of L4-S1 segment (Model A) and a L5 bilateral spondylolysis model (Model B) were established using Mimics, Geomagic, and SolidWorks. Additional models were constructed incorporating translaminar lag screws of varying diameters: 4.5 mm (solid, Model C1; cannulated, Model C2), 4.0 mm (solid, Model D1; cannulated, Model D2), and 3.5 mm (solid, Model E1; cannulated, Model E2). Following model validation, a 400 N axial load and a 10 N m moment were applied to simulate five conditions. Analyzed parameters included range of motion (ROM), maximum displacement, and maximum von Mises stress in the disc, pars interarticularis, and fixation screws.
Results
Compared to the intact model (A), the spondylolysis model (B) showed significant instability. All fixation models (C1-E2) effectively restored ROM to intact levels, regardless of screw diameter or structural design. Larger-diameter (4.5 mm) and solid designs demonstrated slightly better control of model displacement. Increasing diameter significantly reduced stress concentrations within the pars interarticularis and fixation screws. For a given diameter, cannulated screws sustained higher stress than solid screws. The 3.5 mm cannulated screw exhibited the highest screw stress, approaching the yield strength of titanium alloy under rotational loading.
Conclusion
All translaminar lag screw designs can effectively restore stability in L5 spondylolysis. While different diameters and structural designs have a minor impact on stability restoration, they significantly affect implant stress. In this specific patient-derived model, the 4.5 mm solid screw demonstrated the most favorable biomechanical profile, with the lowest stress on the implant and effective control of graft displacement. These findings suggest a potential mechanical advantage for 4.5 mm solid screws; however, confirmation through multi-specimen studies or clinical trials is required before broader clinical recommendations can be made.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Spine Surgery, The 940th Hospital of Joint Logistic Support Force of Chinese People's Liberation Army, Lanzhou, Gansu, China.
- Department of Orthopedics, Tangdu Hospital, Fourth Military Medical University, Xi'an, Shaanxi, China.
- Department of Orthopedics, Shexian Hospital, Handan, Hebei, China.
- First Clinical Medical School, Gansu University of Chinese Medical, Lanzhou, Gansu, China.
Zitieren
APA
Li, F., Tan, X., Zhang, T., Liu, H., & Li, S. (2026). Three-dimensional finite element analysis of translaminar lag screw designs for L5 bilateral spondylolysis. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1882221
Vancouver
Li F, Tan X, Zhang T, Liu H, Li S. Three-dimensional finite element analysis of translaminar lag screw designs for L5 bilateral spondylolysis. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1882221
BibTeX
@article{Li2026,
author = {Li, Feng and Tan, Xingguo and Zhang, Tao and Liu, Hua and Li, Songkai},
title = {Three-dimensional finite element analysis of translaminar lag screw designs for L5 bilateral spondylolysis},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
volume = {14},
doi = {10.3389/fbioe.2026.1882221}
}
EUVIMED