Repeatability and agreement of planar lumbar range of motion measured by dynamic fluoroscopy and optical motion capture in an ovine cadaveric model
Frontiers in Bioengineering and Biotechnology · 10.06.2026 · 14
DOI: 10.3389/fbioe.2026.1827364 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 10.06.2026. Verfasst von Hongtao Li und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Spine Surgery, Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing, China. Die Arbeit nennt 23 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Introduction
Lumbar range of motion is an important biomechanical parameter for understanding spinal function. However, the repeatability, measurement error, and between-method agreement of dynamic fluoroscopy and optical motion capture remain insufficiently characterized in controlled cadaveric spine testing.
Methods
Planar lumbar kinematics were measured using dynamic fluoroscopy within a dynamic X-ray system (DXRS) and optical motion capture with bone-fixed markers (OMCS) in six fresh-frozen ovine lumbar spine specimens, each comprising a continuous L1-L5 segment, during quasi-static flexion-extension and lateral bending. Three repeated trials were performed for each condition. Repeatability was assessed using the intraclass correlation coefficient (ICC(2,1)), the standard error of measurement (SEM), and the minimum detectable change at the 95% confidence level (MDC95). Between-method agreement was evaluated using a repeated-measures Bland-Altman approach with specimen-level clustering, exploratory proportional-bias assessment, and specimen-level bootstrap confidence intervals (CIs) (10,000 replicates) for the bias and limits of agreement (LoA).
Results
ICC point estimates were generally high within each method across motion conditions and vertebral levels; however, given the modest specimen sample (n = 6), the corresponding 95% CIs were wide and these estimates are reported as preliminary, condition-specific values. OMCS showed numerically smaller SEM and MDC95 than DXRS in this dataset, but specimen-level bootstrap intervals indicated uncertainty around most between-method differences, so these descriptive differences should not be interpreted as evidence of general method superiority. Between-method analysis showed a small positive OMCS-minus-DXRS bias with condition-dependent LoA, and these agreement estimates should likewise be interpreted as preliminary and condition-specific.
Conclusion
In this controlled ovine cadaveric model, both DXRS and bone-fixed OMCS provided repeatable measurements of projected planar lumbar motion. OMCS showed numerically smaller measurement-error indices in this dataset and under the present bone-fixed ex vivo ovine configuration; these descriptive, configuration-specific differences do not imply general method superiority, clinical equivalence, or interchangeability with routine human skin-marker motion capture. Cross-device comparisons should account for systematic bias, LoA, and uncertainty related to repeated-measures clustering. The reported error and agreement estimates may help inform method selection and interpretation of longitudinal change in preclinical spine biomechanics studies.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Spine Surgery, Wangjing Hospital, China Academy of Chinese Medical Sciences, Beijing, China.
- Beijing Key Laboratory of Digital Intelligence Traditional Chinese Medicine for Preventing and Treating Degenerative Bone and Joint Diseases, Beijing, China.
- Department of Radiology, Peking Union Medical College Hospital, Chinese Academy of Medical Sciences and Peking Union Medical College, Beijing, China.
Zitieren
APA
Li, H., Liu, G., Han, C., Peng, B., Qin, R., Zheng, J., Yang, G., Zhu, L., Wang, Y., & Feng, M. (2026). Repeatability and agreement of planar lumbar range of motion measured by dynamic fluoroscopy and optical motion capture in an ovine cadaveric model. Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1827364
Vancouver
Li H, Liu G, Han C, Peng B, Qin R, Zheng J, et al. Repeatability and agreement of planar lumbar range of motion measured by dynamic fluoroscopy and optical motion capture in an ovine cadaveric model. Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1827364
BibTeX
@article{Li2026,
author = {Li, Hongtao and Liu, Guangwei and Han, Changxiao and Peng, Bochen and Qin, Ruiyao and Zheng, Jiacheng and Yang, Guangyi and Zhu, Liguo and Wang, Yun and Feng, Minshan},
title = {Repeatability and agreement of planar lumbar range of motion measured by dynamic fluoroscopy and optical motion capture in an ovine cadaveric model},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
volume = {14},
doi = {10.3389/fbioe.2026.1827364}
}
EUVIMED