Comparison of correction formulas for intraocular pressure measured by Goldmann tonometer following various refractive surgeries (FS-LASIK, SMILE, tPRK)
Frontiers in Bioengineering and Biotechnology · 22.01.2026 · 14
DOI: 10.3389/fbioe.2026.1735079 ↗Kurz gefasst
Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 22.01.2026. Verfasst von Jie Tong und 15 weiteren Autor:innen. Beteiligte Einrichtung: National Clinical Research Center for Ocular Diseases, Eye Hospital, WenZhou Medical University, Wenzhou, China. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 46 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
This study aims to compare different intraocular pressure (IOP) correction formulas in post-refractive surgery patients who underwent Femtosecond Laser-Assisted In Situ Keratomileusis (FS-LASIK), Small Incision Lenticule Extraction (SMILE), and Trans-epithelial Photorefractive Keratectomy (tPRK), and to identify the most accurate formula for correcting IOP values measured by Goldmann applanation tonometer (GAT).
Methods
This prospective study included 160 patients (160 eyes) who underwent FS-LASIK (58 eyes), SMILE (52 eyes), or tPRK (50 eyes) at the Eye Hospital, Wenzhou Medical University. IOP was measured using GAT and Dynamic Contour Tonometry (DCT) preoperatively and 3 months postoperatively. Optical zone diameter, ablation depth (AD), residual stromal thickness, refractive error correction, central corneal thickness (CCT), and mean corneal curvature (Km) were collected. Fourteen published GAT correction formulas (F1–F14) were applied to postoperative GAT values and compared with preoperative measurements. Differences between pre- and postoperative DCT readings served as a reference.
Results
In FS-LASIK and SMILE, formulas F3, F4, and F11 showed superior correction performance. In FS-LASIK, the mean differences, concordance correlation coefficients (CCC), and proportions of differences within ±2 mmHg were closer to or better than the DCT reference (0.58, 0.572, 55.17%), with F3 (0.25, 0.372, 72.41%), F4 (0.20, 0.373, 63.79%), and F11 (0.04, 0.324, 67.24%) showing high agreement. Similarly, in SMILE, F3 (0.25, 0.379, 57.69%), F4 (0.22, 0.375, 61.54%), and F11 (0.01, 0.399, 63.46%) outperformed or approximated DCT (0.86, 0.447, 44.23%). In tPRK, F1 (0.18, 0.653, 64.00%) and F2 (0.25, 0.316, 62.00%) provided better correction than DCT (0.19, 0.493, 56.00%).
Conclusion
Formulas F3, F4, and F11 are applicable for GAT IOP correction in both FS-LASIK and SMILE eyes. F1 and F2 are more suitable for tPRK; the preoperative IOP parameters included in F1 may lead to assessment bias, making F2 the recommended choice when preoperative IOP is unavailable.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- National Clinical Research Center for Ocular Diseases, Eye Hospital, WenZhou Medical University, Wenzhou, China.
- State Key Laboratory of Ophthalmology, Optometry and Vision Sicence, Eye Hospital, WenZhou Medical University, Wenzhou, China.
- National Engineering Research Center of Ophthalmology and Optometry, Eye Hospital, WenZhou Medical University, Wenzhou, China.
- ZheJiang-United Kingdom Joint Lab of Ocular Biomechanics, Wenzhou, China.
- The Institute of Ocular Biomechanics, Wenzhou Medical University, Wenzhou, China.
- OuJiang Laboratory, Eye Hospital, WenZhou Medical University, Wenzhou, China.
Zitieren
APA
Tong, J., Zhou, K., Wu, H., Zhao, Y., Gu, Z., Li, Y., Zong, J., Tong, X., Zhou, X., Wang, L., Chen, S., Qu, J., Wang, Q., Xu, D., Elsheikh, A., & Bao, F. (2026). Comparison of correction formulas for intraocular pressure measured by Goldmann tonometer following various refractive surgeries (FS-LASIK, SMILE, tPRK). Frontiers in Bioengineering and Biotechnology, 14. https://doi.org/10.3389/fbioe.2026.1735079
Vancouver
Tong J, Zhou K, Wu H, Zhao Y, Gu Z, Li Y, et al. Comparison of correction formulas for intraocular pressure measured by Goldmann tonometer following various refractive surgeries (FS-LASIK, SMILE, tPRK). Frontiers in Bioengineering and Biotechnology. 2026;14. doi:10.3389/fbioe.2026.1735079
BibTeX
@article{Tong2026,
author = {Tong, Jie and Zhou, Kai and Wu, HongJiang and Zhao, YeWei and Gu, ZhanHao and Li, YiQian and Zong, JiaHui and Tong, XuanYa and Zhou, XiaoFei and Wang, LuMeng and Chen, ShiHao and Qu, Jia and Wang, QinMei and Xu, DongYe and Elsheikh, Ahmed and Bao, FangJun},
title = {Comparison of correction formulas for intraocular pressure measured by Goldmann tonometer following various refractive surgeries (FS-LASIK, SMILE, tPRK)},
journal = {Frontiers in Bioengineering and Biotechnology},
year = {2026},
volume = {14},
doi = {10.3389/fbioe.2026.1735079}
}
EUVIMED