Effects of IMU sensor-to-segment calibration on clinical 3D elbow joint angles estimation

Alessandro Bonfiglio, David Tacconi, Raoul M Bongers, Elisabetta Farella

Frontiers in Bioengineering and Biotechnology · 17.05.2024 · 12

DOI: 10.3389/fbioe.2024.1385750 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 17.05.2024. Verfasst von Alessandro Bonfiglio und 3 weiteren Autor:innen. Beteiligte Einrichtung: Information Engineering and Computer Science Department (DISI), University of Trento, Trento, Italy. Die Arbeit wurde bisher 26-mal zitiert (Crossref) und nennt 38 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Inertial Measurement Units (IMU) require a sensor-to-segment calibration procedure in order to compute anatomically accurate joint angles and, thereby, be employed in healthcare and rehabilitation. Research literature proposes several algorithms to address this issue. However, determining an optimal calibration procedure is challenging due to the large number of variables that affect elbow joint angle accuracy, including 3D joint axis, movement performed, complex anatomy, and notable skin artefacts. Therefore, this paper aims to compare three types of calibration techniques against an optical motion capture reference system during several movement tasks to provide recommendations on the most suitable calibration for the elbow joint.

Methods

Thirteen healthy subjects were instrumented with IMU sensors and optical marker clusters. Each participant performed a series of static poses and movements to calibrate the instruments and, subsequently, performed single-plane and multi-joint tasks. The metrics used to evaluate joint angle accuracy are Range of Motion (ROM) error, Root Mean Squared Error (RMSE), and offset. We performed a three-way RM ANOVA to evaluate the effect of joint axis and movement task on three calibration techniques: N-Pose (NP), Functional Calibration (FC) and Manual Alignment (MA).

Results

Despite small effect sizes in ROM Error, NP displayed the least precision among calibrations due to interquartile ranges as large as 24.6°. RMSE showed significant differences among calibrations and a large effect size where MA performed best (RMSE = 6.3°) and was comparable with FC (RMSE = 7.2°). Offset showed a large effect size in the calibration*axes interaction where FC and MA performed similarly.

Conclusion

Therefore, we recommend MA as the preferred calibration method for the elbow joint due to its simplicity and ease of use. Alternatively, FC can be a valid option when the wearer is unable to hold a predetermined posture.

Schlagwörter

Stichwörter der Autor:innen

IMUcarrying angleelbow biomechanicsjoint angle modellingsensor-to-segment calibration

Autor:innen und Institutionen

  • Information Engineering and Computer Science Department (DISI), University of Trento, Trento, Italy.
  • Euleria Health, Rovereto, Italy.
  • Energy Efficient Embedded Digital Architectures, Fondazione Bruno Kessler (FBK), Trento, Italy.
  • Department of Human Movement Sciences, University Medical Center Groningen, Groningen, Netherlands.

Zitieren

APA

Bonfiglio, A., Tacconi, D., Bongers, R. M., & Farella, E. (2024). Effects of IMU sensor-to-segment calibration on clinical 3D elbow joint angles estimation. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1385750

Vancouver

Bonfiglio A, Tacconi D, Bongers RM, Farella E. Effects of IMU sensor-to-segment calibration on clinical 3D elbow joint angles estimation. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1385750

BibTeX
@article{Bonfiglio2024,
  author = {Bonfiglio, Alessandro and Tacconi, David and Bongers, Raoul M and Farella, Elisabetta},
  title = {Effects of IMU sensor-to-segment calibration on clinical 3D elbow joint angles estimation},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2024},
  volume = {12},
  doi = {10.3389/fbioe.2024.1385750}
}
Kostenlos starten →