EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

In vitro biocompatibility evaluation of functional electrically stimulating microelectrodes on primary glia

Christopher T Tsui, Soroush Mirkiani, David A Roszko, Matthew A Churchward, Vivian K Mushahwar, Kathryn G Todd

Frontiers in Bioengineering and Biotechnology · 19.01.2024 · 12

DOI: 10.3389/fbioe.2024.1351087 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 19.01.2024. Verfasst von Christopher T Tsui und 5 weiteren Autor:innen. Beteiligte Einrichtung: Department of Biomedical Engineering, University of Alberta, Edmonton, AB, Canada. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 91 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Neural interfacing devices interact with the central nervous system to alleviate functional deficits arising from disease or injury. This often entails the use of invasive microelectrode implants that elicit inflammatory responses from glial cells and leads to loss of device function. Previous work focused on improving implant biocompatibility by modifying electrode composition; here, we investigated the direct effects of electrical stimulation on glial cells at the electrode interface. A high-throughput in vitro system that assesses primary glial cell response to biphasic stimulation waveforms at 0 mA, 0.15 mA, and 1.5 mA was developed and optimized. Primary mixed glial cell cultures were generated from heterozygous CX3CR-1 +/EGFP mice, electrically stimulated for 4 h/day over 3 days using 75 μm platinum-iridium microelectrodes, and biomarker immunofluorescence was measured. Electrodes were then imaged on a scanning electron microscope to assess sustained electrode damage. Fluorescence and electron microscopy analyses suggest varying degrees of localized responses for each biomarker assayed (Hoescht, EGFP, GFAP, and IL-1β), a result that expands on comparable in vivo models. This system allows for the comparison of a breadth of electrical stimulation parameters, and opens another avenue through which neural interfacing device developers can improve biocompatibility and longevity of electrodes in tissue.

Schlagwörter

Stichwörter der Autor:innen

astrocytebiocompatibilityelectrical stimulationmicroelectrodemicroglianeuroinflammation

Autor:innen und Institutionen

  • Department of Biomedical Engineering, University of Alberta, Edmonton, AB, Canada.
  • Neurochemical Research Unit, Department of Psychiatry, University of Alberta, Edmonton, AB, Canada.
  • Neuroscience and Mental Health Institute (NMHI), University of Alberta, Edmonton, AB, Canada.
  • Sensory Motor Adaptive Rehabilitation Technology (SMART) Network, University of Alberta, Edmonton, AB, Canada.
  • Department of Biological and Environmental Sciences, Concordia University of Edmonton, Edmonton, AB, Canada.
  • Division of Physical Medicine and Rehabilitation, Department of Medicine, University of Alberta, Edmonton, AB, Canada.

Zitieren

APA

Tsui, C. T., Mirkiani, S., Roszko, D. A., Churchward, M. A., Mushahwar, V. K., & Todd, K. G. (2024). In vitro biocompatibility evaluation of functional electrically stimulating microelectrodes on primary glia. Frontiers in Bioengineering and Biotechnology, 12. https://doi.org/10.3389/fbioe.2024.1351087

Vancouver

Tsui CT, Mirkiani S, Roszko DA, Churchward MA, Mushahwar VK, Todd KG. In vitro biocompatibility evaluation of functional electrically stimulating microelectrodes on primary glia. Frontiers in Bioengineering and Biotechnology. 2024;12. doi:10.3389/fbioe.2024.1351087

BibTeX
@article{Tsui2024,
  author = {Tsui, Christopher T and Mirkiani, Soroush and Roszko, David A and Churchward, Matthew A and Mushahwar, Vivian K and Todd, Kathryn G},
  title = {In vitro biocompatibility evaluation of functional electrically stimulating microelectrodes on primary glia},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2024},
  volume = {12},
  doi = {10.3389/fbioe.2024.1351087}
}
Kostenlos starten →