EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations

Antoine Vautrin, Jensen Aw, Ed Attenborough, Peter Varga

Frontiers in Bioengineering and Biotechnology · 10.08.2023 · 11

DOI: 10.3389/fbioe.2023.1240125 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 10.08.2023. Verfasst von Antoine Vautrin und 3 weiteren Autor:innen. Beteiligte Einrichtung: AO Research Institute Davos, Davos, Switzerland. Die Arbeit wurde bisher 10-mal zitiert (Crossref) und nennt 43 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Porous dental implants represent a promising strategy to reduce failure rate by favoring osseointegration or delivering drugs locally. Incorporating porous features weakens the mechanical capacity of an implant, but sufficient fatigue strength must be ensured as regulated in the ISO 14801 standard. Experimental fatigue testing is a costly and time-intensive part of the implant development process that could be accelerated with validated computer simulations. This study aimed at developing, calibrating, and validating a numerical workflow to predict fatigue strength on six porous configurations of a simplified implant geometry.

Methods

Mechanical testing was performed on 3D-printed titanium samples to establish a direct link between endurance limit (i.e., infinite fatigue life) and monotonic load to failure, and a finite element model was developed and calibrated to predict the latter. The tool was then validated by predicting the fatigue life of a given porous configuration.

Results

The normalized endurance limit (10% of the ultimate load) was the same for all six porous designs, indicating that monotonic testing was a good surrogate for endurance limit. The geometry input of the simulations influenced greatly their accuracy. Utilizing the as-designed model resulted in the highest prediction error (23%) and low correlation between the estimated and experimental loads to failure (R 2 = 0.65). The prediction error was smaller when utilizing specimen geometry based on micro computed tomography scans (14%) or design models adjusted to match the printed porosity (8%).

Discussion

The validated numerical workflow presented in this study could therefore be used to quantitatively predict the fatigue life of a porous implant, provided that the effect of manufacturing on implant geometry is accounted for.

Schlagwörter

Stichwörter der Autor:innen

additive manufacturingdental implantfatiguefinite element analysismicro-CTporous titanium

Autor:innen und Institutionen

  • AO Research Institute Davos, Davos, Switzerland.
  • Graduate School for Cellular and Biomedical Sciences, University of Bern, Bern, Switzerland.
  • Attenborough Dental Laboratories Ltd, Nottingham, United Kingdom.

Zitieren

APA

Vautrin, A., Aw, J., Attenborough, E., & Varga, P. (2023). Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1240125

Vancouver

Vautrin A, Aw J, Attenborough E, Varga P. Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1240125

BibTeX
@article{Vautrin2023,
  author = {Vautrin, Antoine and Aw, Jensen and Attenborough, Ed and Varga, Peter},
  title = {Fatigue life of 3D-printed porous titanium dental implants predicted by validated finite element simulations},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2023},
  volume = {11},
  doi = {10.3389/fbioe.2023.1240125}
}
Kostenlos starten →