Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 23.05.2023. Verfasst von Dongwoo Hahn und 8 weiteren Autor:innen. Beteiligte Einrichtung: Division of Cardiology, Department of Internal Medicine, Seoul National University Bundang Hospital, Seongnam, Republic of Korea. Die Arbeit wurde bisher 8-mal zitiert (Crossref) und nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

The ultrathin-strut drug-eluting stent (DES) has shown better clinical results than thin- or thick-strut DES. We investigated if re-endothelialization was different among three types of DES: ultrathin-strut abluminal polymer-coated sirolimus-eluting stent (SES), thin-strut circumferential polymer-coated everolimus-eluting stent (EES), and thick-strut polymer-free biolimus-eluting stent (BES) to gain insight into the effect of stent design on promoting vascular healing.

Methods

After implanting three types of DES in the coronary arteries of minipigs, we performed optical coherence tomography (OCT) at weeks 2, 4, and 12 ( n = 4, each). Afterward, we harvested the coronary arteries and performed immunofluorescence for endothelial cells (ECs), smooth muscle cells (SMCs), and nuclei. We obtained 3D stack images of the vessel wall and reconstructed the en face view of the inner lumen. We compared re-endothelialization and associated factors among the different types of stents at different time points.

Results

SES showed significantly faster and denser re-endothelialization than EES and BES at weeks 2 and 12. Especially in week 2, SES elicited the fastest SMC coverage and greater neointimal cross-sectional area (CSA) compared to EES and BES. A strong correlation between re-endothelialization and SMC coverage was observed in week 2. However, the three stents did not show any difference at weeks 4 and 12 in SMC coverage and neointimal CSA. At weeks 2 and 4, SMC layer morphology showed a significant difference between stents. A sparse SMC layer was associated with denser re-endothelialization and was significantly higher in SES. Unlike the sparse SMC layer, the dense SMC layer did not promote re-endothelialization during the study period.

Conclusion

Re-endothelialization after stent implantation was related to SMC coverage and SMC layer differentiation, which were faster in SES. Further investigation is needed to characterize the differences among the SMCs and explore methods for increasing the sparse SMC layer in order to improve stent design and enhance safety and efficacy.

Schlagwörter

Stichwörter der Autor:innen

drug-eluting stent (DES)neointimal coverageoptical coherence tomographyre-endothelializationsmooth muscle cell (SMC)

Autor:innen und Institutionen

  • Division of Cardiology, Department of Internal Medicine, Seoul National University Bundang Hospital, Seongnam, Republic of Korea.
  • Cardiovascular center, Asan Chungmu Hospital, Asan, Republic of Korea.
  • College of Medicine, Seoul National University, Seoul, Republic of Korea.

Zitieren

APA

Hahn, D., Lee, D., Hyun, W., Cho, Y., Yoon, C. H., Jeon, K. H., Kang, S. H., Youn, T. J., & Chae, I. H. (2023). Faster smooth muscle cell coverage in ultrathin-strut drug-eluting stent leads to earlier re-endothelialization. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1207858

Vancouver

Hahn D, Lee D, Hyun W, Cho Y, Yoon CH, Jeon KH, et al. Faster smooth muscle cell coverage in ultrathin-strut drug-eluting stent leads to earlier re-endothelialization. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1207858

BibTeX
@article{Hahn2023,
  author = {Hahn, Dongwoo and Lee, Donghoon and Hyun, Woonggyu and Cho, Yunnie and Yoon, Chang-Hwan and Jeon, Ki-Hyun and Kang, Si-Hyuck and Youn, Tae-Jin and Chae, In-Ho},
  title = {Faster smooth muscle cell coverage in ultrathin-strut drug-eluting stent leads to earlier re-endothelialization},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2023},
  volume = {11},
  doi = {10.3389/fbioe.2023.1207858}
}
Kostenlos starten →