Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: A finite element analysis

Qiang Huang, CongMing Zhang, HuanAn Bai, Qian Wang, Zhong Li, Yao Lu, Teng Ma

Frontiers in Bioengineering and Biotechnology · 21.02.2023 · 11

DOI: 10.3389/fbioe.2023.1116976 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Frontiers in Bioengineering and Biotechnology, online erschienen am 21.02.2023. Verfasst von Qiang Huang und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of Orthopedics, Hong Hui Hospital, Xi'an Jiaotong University, Xi'an, Shaanxi, China. Die Arbeit wurde bisher 34-mal zitiert (Crossref) und nennt 37 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

The existing implants for fixation of femoral neck fractures have poor biomechanical stability, so the failure rate is high. We designed two modified intramedullary implants for treating unstable femoral neck fractures (UFNFs). We tried to improve the biomechanical stability of fixation by shortening the moment and reducing stress concentration. Each modified intramedullary implant was compared with cannulated screws (CSs) through finite element analysis (FEA).

Methods

Five different models were included: three cannulated screws (CSs, Model 1) in an inverted triangle configuration, the dynamic hip screw with an anti-rotation screw (DHS + AS, Model 2), the femoral neck system (FNS, Model 3), the modified intramedullary femoral neck system (IFNS, Model 4), and the modified intramedullary interlocking system (IIS, Model 5). Three-dimensional (3D) models of femur and implants were constructed by using 3D modelling software. Three load cases were simulated to assess the maximal displacement of models and fracture surface. The maximal stress at the bone and implants was also evaluated.

Results

FEA data showed that Model 5 had the best performance in terms of maximum displacement while Model 1 had the worst performance for this index under axial load of 2100 N. With respect to Maximum stress, Model 4 had the best performance while Model 2 had the worst performance under axial load. The general trends under bending and torsion load were consistent with that under axial load. Our data demonstrated that the two modified intramedullary implants exhibited the best biomechanical stability, followed by FNS and DHS + AS, and then three cannulated screws in axial, bending, and torsion load cases.

Conclusion

The two modified intramedullary designs showed the best biomechanical performance among the five implants included in this study. Therefore, this might provide some new options for trauma surgeons to deal with unstable femoral neck fractures.

Schlagwörter

Stichwörter der Autor:innen

axialbendingbiomechanicalfemoral neck fracturefinite element modelintramedullarytorsion

Autor:innen und Institutionen

  • Department of Orthopedics, Hong Hui Hospital, Xi'an Jiaotong University, Xi'an, Shaanxi, China.

Zitieren

APA

Huang, Q., Zhang, C., Bai, H., Wang, Q., Li, Z., Lu, Y., & Ma, T. (2023). Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: A finite element analysis. Frontiers in Bioengineering and Biotechnology, 11. https://doi.org/10.3389/fbioe.2023.1116976

Vancouver

Huang Q, Zhang C, Bai H, Wang Q, Li Z, Lu Y, et al. Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: A finite element analysis. Frontiers in Bioengineering and Biotechnology. 2023;11. doi:10.3389/fbioe.2023.1116976

BibTeX
@article{Huang2023,
  author = {Huang, Qiang and Zhang, CongMing and Bai, HuanAn and Wang, Qian and Li, Zhong and Lu, Yao and Ma, Teng},
  title = {Biomechanical evaluation of two modified intramedullary fixation system for treating unstable femoral neck fractures: A finite element analysis},
  journal = {Frontiers in Bioengineering and Biotechnology},
  year = {2023},
  volume = {11},
  doi = {10.3389/fbioe.2023.1116976}
}
Kostenlos starten →