EUVIMEDEUVIMED entdecken
VergleichsstudieOpen Access · CC BY-NC 4.0

Comparison of Toric Intraocular Lens Axis Accuracy between Optical Biometry and Dual Scheimpflug Topography

Seonghwan Kim, Yengwoo Son, Joon Young Hyon

Korean Journal of Ophthalmology · 2025

DOI: 10.3341/kjo.2024.0117 ↗

Kurz gefasst

Vergleichsstudie in der Zeitschrift Korean Journal of Ophthalmology, 2025. Verfasst von Seonghwan Kim und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Ophthalmology, Seoul Metropolitan Government Seoul National University Boramae Medical Center, Seoul, Korea. Hauptthemen laut MEDLINE: Biometry, Lenses, Intraocular, Corneal Topography, Astigmatism. Die Arbeit nennt 25 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0). Gefördert von: Seoul National University Bundang Hospital.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

To evaluate the accuracy of toric intraocular lens (IOL) axis prediction between two preoperative measurement devices: the optical biometry (IOLMaster 500 or IOLMaster 700) and the dual Scheimpflug topography (Galilei G4)

Methods

Medical records of 64 eyes from 44 patients who underwent phacoemulsification and posterior chamber toric IOL (Zeiss AT TORBI 709M) implantation between July 2017 and January 2022 were reviewed. All patients underwent preoperative evaluation by optical biometry (IOLMaster 500 or IOLMaster 700) and Galilei G4. The gold-standard axis that minimizes astigmatism was calculated by the online Toric Results Analyzer postoperatively and compared to the preoperative toric IOL axis calculated by the Z CALC Online IOL Calculator using parameters from either IOLMaster or Galilei G4. The axis error (AE) and the absolute AE (AAE) between the gold-standard axis and the preoperative calculated axis were analyzed to assess the accuracy of each device.

Results

The mean flat keratometry and steep keratometry were 42.99 diopters (D) and 45.61 D, respectively, in IOLMaster, and 43.04 D and 45.51 D, respectively, in Galilei G4, which did not show any significant difference. The mean keratometric astigmatism was 2.62 D in IOLMaster and 2.46 D in Galilei G4, which also did not show any statistical difference. The keratometric astigmatism axis did not show any significant difference between IOLMaster and Galilei G4. The mean AE and AAE were 0.19° and 6.84°, respectively, by IOLMaster, and –0.80° and 7.98°, respectively, by Galilei G4. The AE and AAE by IOLMaster did not show any significant difference compared to those of Galilei G4 (p = 0.583, and p = 0.346, respectively).

Conclusions

This study suggests that the Galilei G4 demonstrated a similar level of accuracy to the IOLMaster in predicting the toric IOL axis, based on the gold-standard axis provided by the Toric Results Analyzer.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BiometryLenses, IntraocularCorneal TopographyAstigmatismRefraction, OcularPhacoemulsificationLens Implantation, IntraocularHumansFemaleMaleAgedMiddle AgedRetrospective StudiesReproducibility of ResultsVisual AcuityAged, 80 and over

Stichwörter der Autor:innen

Dual Scheimpflug analyzerOptical biometryToric Results AnalyzerToric intraocular lens axis

Autor:innen und Institutionen

  • Department of Ophthalmology, Seoul Metropolitan Government Seoul National University Boramae Medical Center, Seoul, Korea.
  • Institute of Vision Research, Department of Ophthalmology, Severance Hospital, Yonsei University College of Medicine, Seoul, Korea.
  • Department of Ophthalmology, Seoul National University Bundang Hospital, Seongnam, Korea.

Förderung

  • Seoul National University Bundang Hospital

Zitieren

APA

Kim, S., Son, Y., & Hyon, J. Y. (2025). Comparison of Toric Intraocular Lens Axis Accuracy between Optical Biometry and Dual Scheimpflug Topography. Korean Journal of Ophthalmology. https://doi.org/10.3341/kjo.2024.0117

Vancouver

Kim S, Son Y, Hyon JY. Comparison of Toric Intraocular Lens Axis Accuracy between Optical Biometry and Dual Scheimpflug Topography. Korean Journal of Ophthalmology. 2025. doi:10.3341/kjo.2024.0117

BibTeX
@article{Kim2025,
  author = {Kim, Seonghwan and Son, Yengwoo and Hyon, Joon Young},
  title = {Comparison of Toric Intraocular Lens Axis Accuracy between Optical Biometry and Dual Scheimpflug Topography},
  journal = {Korean Journal of Ophthalmology},
  year = {2025},
  doi = {10.3341/kjo.2024.0117}
}