Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung
Archives of Pediatric Critical Care · 30.06.2026 · 4 (1) · 9-16
DOI: 10.32990/apcc.2026.00038 ↗Kurz gefasst
Fachartikel in der Zeitschrift Archives of Pediatric Critical Care, erschienen am 30.06.2026. Verfasst von Weerapong Lilitwat und 1 weiteren Autor:innen. Die Arbeit wurde bisher 3-mal zitiert (Crossref) und nennt 14 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Contemporary pediatric high-frequency oscillatory ventilators may differ in how set oscillatory parameters are translated into delivered tidal volume under varying respiratory mechanical loads.We evaluated determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung using a device-specific full-factorial bench model.
Methods
A Fabian high-frequency oscillatory oscillator was connected to an IngMar pediatric mechanical test lung through a standard dry pediatric circuit.Tidal volume was measured with a Florian respiratory monitor placed between the circuit and lung model.We tested 1,512 combinations of frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, compliance, and resistance.Measurements were recorded after at least 2 minutes of steady state at each condition.Analysis of variance and multiple linear regression were used to evaluate associations with delivered tidal volume.
Results
Delivered tidal volume was associated with frequency, pressure amplitude, mean airway pressure, inspiratory-to-expiratory ratio, and resistance (all P
Zitieren
APA
Lilitwat, W., & Vorakunthada, Y. (2026). Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung. Archives of Pediatric Critical Care, 4(1), 9-16. https://doi.org/10.32990/apcc.2026.00038
Vancouver
Lilitwat W, Vorakunthada Y. Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung. Archives of Pediatric Critical Care. 2026;4(1):9-16. doi:10.32990/apcc.2026.00038
BibTeX
@article{Lilitwat2026,
author = {Lilitwat, Weerapong and Vorakunthada, Yuttiwat},
title = {Determinants of delivered tidal volume during high-frequency oscillatory ventilation in a pediatric mechanical test lung},
journal = {Archives of Pediatric Critical Care},
year = {2026},
volume = {4},
number = {1},
pages = {9--16},
doi = {10.32990/apcc.2026.00038}
}
EUVIMED