Effects of flow rate accuracy in two-day anticancer drug infusion with disposable pumps on plasma drug concentrations

Kyoung Jin Lee, Jung Chan Lee

Technology and Health Care · 2024

DOI: 10.3233/thc-230227 ↗

Kurz gefasst

Fachartikel in der Zeitschrift Technology and Health Care, 2024. Verfasst von Kyoung Jin Lee und 1 weiteren Autor:innen. Beteiligte Einrichtung: Interdisciplinary Program in Bioengineering, Seoul National University Graduate School, Seoul, Korea. Hauptthemen laut MEDLINE: Fluorouracil, Infusion Pumps. Die Arbeit nennt 23 Literaturangaben.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND: Elastomeric pumps have a curved infusion rate profile over infusion time. Chemically driven pumps can overcome such limitations of elastomeric pumps and infuse constantly. However, studies on the pharmacokinetic benefit of chemically-driven pumps are insufficient. OBJECTIVE: This study aimed to determine effects of constant infusion with a chemically-driven pump on plasma drug concentrations compared to elastomeric pumps. METHODS: Infusion rate profiles of a chemically driven pump and two elastomeric pumps were measured in vitro tests under three height conditions of drug reservoir. Plasma drug concentrations were estimated using a pharmacokinetic model of 5-fluorouracil (5FU). RESULTS: The chemically-driven pump was more accurate than elastomeric pumps during the total infusion time (Root-mean-square-error (RMSE): 3% vs. 13%) which thus reduced its deviation of plasma 5FU concentration over time to one-fifth of that with an elastomeric pump. The chemically-driven pump had less than 5% of RMSE despite the influence of height difference. CONCLUSION: Although chemically-driven pumps maintained plasma 5FU concentration successfully and elastomeric pumps did not, both pumps were proper for 5FU infusion because the time-dependent changes in infusion rate did not affect the area under the curve. Chemically driven pumps would be more advantageous for drugs that are sensitive to their plasma concentrations.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

FluorouracilInfusion PumpsHumansAntineoplastic AgentsInfusions, IntravenousElastomersEquipment DesignDisposable Equipment

Stichwörter der Autor:innen

5-fluorouracilInfusion pumpsdisposable equipmentpharmacokinetics

Substanzen

FluorouracilAntineoplastic AgentsElastomers

Autor:innen und Institutionen

  • Interdisciplinary Program in Bioengineering, Seoul National University Graduate School, Seoul, Korea.
  • Department of Biomedical Engineering, Seoul National University College of Medicine, Seoul, Korea.
  • Institute of Medical and Biological Engineering, Medical Research Center, Seoul National University, Seoul, Korea.
  • Department of Biomedical Engineering and Innovative Medical Technology Research Institute, Seoul National University Hospital, Korea.

Zitieren

APA

Lee, K. J., & Lee, J. C. (2024). Effects of flow rate accuracy in two-day anticancer drug infusion with disposable pumps on plasma drug concentrations. Technology and Health Care. https://doi.org/10.3233/thc-230227

Vancouver

Lee KJ, Lee JC. Effects of flow rate accuracy in two-day anticancer drug infusion with disposable pumps on plasma drug concentrations. Technology and Health Care. 2024. doi:10.3233/thc-230227

BibTeX
@article{Lee2024,
  author = {Lee, Kyoung Jin and Lee, Jung Chan},
  title = {Effects of flow rate accuracy in two-day anticancer drug infusion with disposable pumps on plasma drug concentrations},
  journal = {Technology and Health Care},
  year = {2024},
  doi = {10.3233/thc-230227}
}
Kostenlos starten →