Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?
The Journal of Vascular Access · 2025
DOI: 10.1177/11297298251314615 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift The Journal of Vascular Access, 2025. Verfasst von Arthur Muller und 7 weiteren Autor:innen. Beteiligte Einrichtung: EXPLOR! Center, Air Liquide Healthcare, Bagneux, France. Hauptthemen laut MEDLINE: Anti-Bacterial Agents, Infusion Pumps, Elastomers, Piperacillin, Tazobactam Drug Combination. Die Arbeit nennt 18 Literaturangaben. Gefördert von: Air Liquide Healthcare.
Abstract (Originalsprache)
Background
Outpatient Parenteral Antibiotic Therapy (OPAT) services have expanded throughout the world thanks to elastomeric pumps (EP). Their flowrate is known to vary by ±15%. Treatment effectiveness and the organization of care at home may be impacted, especially with piperacillin/tazobactam infusion (TAZ). An in vitro study was conducted, following the guidelines of EP and drug manufacturers, to assess the impact of three parameters on the accuracy of the infusion flowrate: composition, volume, and temperature of the infused solution.
Methods
A test bench was created to evaluate the accuracy of flowrate infusion when changing one parameter at a time (addition of TAZ in the infused solution, minimal filling volume and flow regulator at ambient temperature) compared to the standard conditions provided by the manufacturer. Two types of EP were used and connected to a weighing scale. Instantaneous and average mass flowrates over the total infusion time were determined from the reservoir mass loss. These values were compared with the nominal flowrates announced by the manufacturers to evaluate the performance of EP under the different conditions. The time the flowrate was within the expected ±15% around the nominal flowrate was calculated as the time in range (TIR).
Results
Deviations of the measured mean flowrate from the nominal flowrate range from −37% to 18% depending on the condition. TIR went from 0% to 80% of the total infusion duration. The test bench showed a reliability of more than 96%.
Conclusions
The test bench highlights differences in performance. These results raise several questions about the sensitivity of EPs for TAZ infusion in OPAT, the norm’s requirements and the need to prescribe a medical device whose limits are well known, as well as therapeutic education or the course of care for treatment administration.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- EXPLOR! Center, Air Liquide Healthcare, Bagneux, France.
- Paris-Saclay University, Laboratoire Complexité, innovations, activités motrices et sportives (CIAMS), Orsay, France.
- Orléans University, CIAMS, Orléans, France.
- Univ. Lille, CHU Lille, ULR 7365 - GRITA - Groupe de Recherche sur les formes Injectables et les Technologies Associées, Lille, France.
- Hôpital Universitaire Cochin, Assistance Publique - Hôpitaux de Paris, pulmonology unit, France.
- Vitalaire, Air Liquide Healthcare, Bagneux, France.
ORCID-Profile: Clara Virbel-Fleischman
Förderung
- Air Liquide Healthcare
Zitieren
APA
Muller, A., Rétory, Y., Négrier, L., Lemarié, B., Pfeifer, A., Schmidt, A., Dupont, C., & Virbel-Fleischman, C. (2025). Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?. The Journal of Vascular Access. https://doi.org/10.1177/11297298251314615
Vancouver
Muller A, Rétory Y, Négrier L, Lemarié B, Pfeifer A, Schmidt A, et al. Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?. The Journal of Vascular Access. 2025. doi:10.1177/11297298251314615
BibTeX
@article{Muller2025,
author = {Muller, Arthur and Rétory, Yann and Négrier, Laura and Lemarié, Benoit and Pfeifer, Anne and Schmidt, Alexandra and Dupont, Christian and Virbel-Fleischman, Clara},
title = {Outpatient parenteral antibiotic therapy with elastomeric pumps, what is the flowrate accuracy?},
journal = {The Journal of Vascular Access},
year = {2025},
doi = {10.1177/11297298251314615}
}
EUVIMED