Validation of substances used for calibrating the Periotron 8000® instrument and conversion equations: an in vitro study
International Journal of Medical Device and Adjuvant Treatments · 2021-10 · 4
DOI: 10.32113/ijmdat_202110_364 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Medical Device and Adjuvant Treatments, 2021-10. Verfasst von M. Jordán-López und 3 weiteren Autor:innen.
Abstract (Originalsprache)
Objective
Periotron 8000® is an electronic instrument that quantifies the volume of gingival crevicular fluid and saliva. Previous literature has proposed use of different fluids for calibration of the instrument, including human serum. The objectives were to compare different fluids used for instrument calibration, determine the correlations, and the most appropriate conversion equation for the model. Materials and Methods: The fluids evaluated were physiological saline, human serum, fetal bovine serum, and saliva.The Periotron 8000® instrument was calibrated with each fluid, and the correlation between these substances was analyzed. The calibration data were adjusted to a straight line, a second-, third-, and fourth-degree polynomial. R2(goodness-of-fit) values and the root mean square error (RMSE) were calculated for each regression model.
Results
All the correlations were significant. However, saliva correlated more strongly with physiological saline solution. The fourth-degree polynomial was the most accurate as a conversion equation because it presented higher R2and lower RMSE.
Conclusions
The four fluids evaluated are useful to calibrate the Periotron 8000® instrument because they produce accurate regression models. Using saliva as a reference, the best fluid for calibration is physiological saline solution.
Schlagwörter
Zitieren
APA
Jordán-López, M., Almiñana-Pastor, P., Alpiste-Illueca, F., & López-Roldán, A. (2021). Validation of substances used for calibrating the Periotron 8000® instrument and conversion equations: an in vitro study. International Journal of Medical Device and Adjuvant Treatments, 4. https://doi.org/10.32113/ijmdat_202110_364
Vancouver
Jordán-López M, Almiñana-Pastor P, Alpiste-Illueca F, López-Roldán A. Validation of substances used for calibrating the Periotron 8000® instrument and conversion equations: an in vitro study. International Journal of Medical Device and Adjuvant Treatments. 2021;4. doi:10.32113/ijmdat_202110_364
BibTeX
@article{JordnLpez2021,
author = {Jordán-López, M. and Almiñana-Pastor, P. and Alpiste-Illueca, F. and López-Roldán, A.},
title = {Validation of substances used for calibrating the Periotron 8000® instrument and conversion equations: an in vitro study},
journal = {International Journal of Medical Device and Adjuvant Treatments},
year = {2021},
volume = {4},
doi = {10.32113/ijmdat_202110_364}
}
EUVIMED