EUVIMEDEUVIMED entdecken
Open Access

Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience

Aakanksha Sriwastwa, Yasmin N Aziz, Kara Weiss, Robert Buse, Bin Zhang iD, Stacie L Demel, Arafat Ali, Sriharsha Voleti, Lily Li-Li Wang, Achala S Vagal iD

American Journal of Neuroradiology · 2025-03 · 46 (3) · 476-482

DOI: 10.3174/ajnr.a8509 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Neuroradiology, online erschienen am 26.09.2024. Verfasst von Aakanksha Sriwastwa und 9 weiteren Autor:innen. Beteiligte Einrichtung: From the Department of Radiology (A.S., L.L.-L.W., A.S.V.), University of Cincinnati Medical Center, Cincinnati, Ohio sriwasaa@ucmail.uc.edu. Hauptthemen laut MEDLINE: Algorithms, Ischemic Stroke, Computed Tomography Angiography, Cerebral Angiography. Die Arbeit wurde bisher 17-mal zitiert (Crossref) und nennt 19 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

BACKGROUND AND PURPOSE: Fast, accurate detection of large (LVO) and medium vessel occlusion (MeVO) is critical for triage and management of acute ischemic stroke. Multiple artificial intelligence (AI)-based software programs are available commercially for automated detection and rapid prioritization of LVO. However, their ability, strengths, and limitations for detection of acute vessel occlusion in the context of expanding indications for mechanical thrombectomy are not entirely understood. We aimed to investigate the performance of a fully automated commercial detection algorithm to detect large and medium vessel occlusions in code stroke patients. MATERIALS AND METHODS: We utilized a single-center, institutional, retrospective registry of all consecutive code stroke patients with CTA and automated processing by using Viz.ai presenting at a large comprehensive stroke center between March 2020 and February 2023. LVO was categorized as anterior LVO (aLVO), defined as occlusion of the intracranial ICA or M1-MCA, and posterior LVO (pLVO), defined as occlusion of the basilar artery or V4-vertebral artery. MeVO was defined as occlusion of the M2-MCA, A1/A2-anterior cerebral artery, or P1/P2-posterior cerebral artery. We compared the accuracy of Viz.ai by using STARD guidelines. Radiology reports from 12 board-certified radiologists were considered the reference standard. Our primary outcome was assessing accuracy of the automated software for aLVO. Our secondary outcome was assessing accuracy for 3 additional categories: all LVO (aLVO and pLVO), aLVO with M2-MCA, and aLVO with MeVO. RESULTS: Of 3590 code stroke patients, 3576 were technically sufficient for analysis by the automated software (median age 67 years; 51% women; 68% white), of which 616 (17.2%) had vessel occlusions. The respective sensitivity and specificity for our prespecified categories were: aLVO: 91% (87-94%), 93% (92-94%); all LVO: 73% (68-77%), 92% (91-93%); aLVO + M2-MCA occlusion: 74% (70-78%), 93% (92-94%); and aLVO + all MeVO: 65% (61-69%), 93% (92-94%). CONCLUSIONS: The automated algorithm demonstrated high accuracy in identifying anterior LVO with lower performance for pLVO and MeVO. It is crucial for acute stroke teams to be aware of the discordance between automated algorithm results and true rates of LVO and MeVO for timely diagnosis and triage.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

AlgorithmsIschemic StrokeComputed Tomography AngiographyCerebral AngiographyHumansRetrospective StudiesFemaleMaleAgedMiddle AgedAged, 80 and overArtificial IntelligenceRegistries

Autor:innen und Institutionen

  • From the Department of Radiology (A.S., L.L.-L.W., A.S.V.), University of Cincinnati Medical Center, Cincinnati, Ohio sriwasaa@ucmail.uc.edu.
  • Department of Neurology and Rehabilitation Medicine (Y.N.A., S.L.D.), University of Cincinnati Medical Center, Cincinnati, Ohio.
  • University of Cincinnati College of Medicine (K.W., R.B., B.Z.), Cincinnati, Ohio.
  • Division of Biostatistics (B.Z.), Cincinnati Children's Hospital Medical Center, Cincinnati, Ohio.
  • Department of Diagnostic Radiology (A.A.), Henry Ford Medical Center, Ann Arbor, Michigan.
  • Department of Diagnostic Radiology (S.V.), University of Pennsylvania, Philadelphia, Pennsylvania.

ORCID-Profile: Bin Zhang, Achala S Vagal

Zitieren

APA

Sriwastwa, A., Aziz, Y. N., Weiss, K., Buse, R., Zhang, B., Demel, S. L., Ali, A., Voleti, S., Wang, L. L., & Vagal, A. S. (2025). Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience. American Journal of Neuroradiology, 46(3), 476-482. https://doi.org/10.3174/ajnr.a8509

Vancouver

Sriwastwa A, Aziz YN, Weiss K, Buse R, Zhang B, Demel SL, et al. Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience. American Journal of Neuroradiology. 2025;46(3):476-482. doi:10.3174/ajnr.a8509

BibTeX
@article{Sriwastwa2025,
  author = {Sriwastwa, Aakanksha and Aziz, Yasmin N and Weiss, Kara and Buse, Robert and Zhang, Bin and Demel, Stacie L and Ali, Arafat and Voleti, Sriharsha and Wang, Lily Li-Li and Vagal, Achala S},
  title = {Performance of Automated Algorithm in Large and Medium Vessel Occlusion Detection: A Real-World Experience},
  journal = {American Journal of Neuroradiology},
  year = {2025},
  volume = {46},
  number = {3},
  pages = {476--482},
  doi = {10.3174/ajnr.a8509}
}
Kostenlos starten →