VOLUME COMPUTED TOMOGRAPHY DOSE INDEX EVALUATION FOR CT SCANNERS OPERATING IN VARIOUS RADIOLOGY DEPARTMENTS ACROSS ALBANIA

Luljeta Disha, Irena Mucollari, Florinda Cfarku

International Journal of Ecosystems and Ecology Science (IJEES) · 02.06.2026 · 16 (3) · 27-32

DOI: 10.31407/ijees16.304 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Ecosystems and Ecology Science (IJEES), online erschienen am 02.06.2026. Verfasst von Luljeta Disha und 2 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Computed Tomography (CT) is widely recognized as a relatively high-dose imaging modality.Therefore, dose optimization and the implementation of appropriate dose constraints are of particular importance.In Albania, the measurement of Volume Computed Tomography Dose Index (CTDIvol) is a mandatory component of quality control (QC) testing for CT scanners, ensuring that CT systems operate within safe radiation dose limits while maintaining optimal image quality for accurate diagnoses.Evaluation of CTDIvol is performed using a pencil ionization chamber in combination with standard PMMA phantoms: a 16cm diameter head phantom and a 32cm diameter body phantom.The aim of this study was to assess CTDIvol values across different types of CT scanners operating in various radiology departments throughout Albania.All data were collected during the initial performance of QC testing procedures.Analysis of the results demonstrated that CTDIvol measurements for the body phantom ranged from 12.5 mGy to 23.4 mGy, while measurements for the head phantom ranged from 35 mGy to 61.3 mGy.The relative differences between experimentally measured CTDIvol values and the corresponding console-displayed values ranged from 0.7% to 17.0% for the body phantom and from 0.3% to 7.6% for the head phantom.This survey provides a valuable baseline assessment of CT scanner's performance across the country.Establishing such reference data enables standardized comparison over time, supports the tracking of equipment performance, and contributes to the optimization of imaging protocols.

Zitieren

APA

Disha, L., Mucollari, I., & Cfarku, F. (2026). VOLUME COMPUTED TOMOGRAPHY DOSE INDEX EVALUATION FOR CT SCANNERS OPERATING IN VARIOUS RADIOLOGY DEPARTMENTS ACROSS ALBANIA. International Journal of Ecosystems and Ecology Science (IJEES), 16(3), 27-32. https://doi.org/10.31407/ijees16.304

Vancouver

Disha L, Mucollari I, Cfarku F. VOLUME COMPUTED TOMOGRAPHY DOSE INDEX EVALUATION FOR CT SCANNERS OPERATING IN VARIOUS RADIOLOGY DEPARTMENTS ACROSS ALBANIA. International Journal of Ecosystems and Ecology Science (IJEES). 2026;16(3):27-32. doi:10.31407/ijees16.304

BibTeX
@article{Disha2026,
  author = {Disha, Luljeta and Mucollari, Irena and Cfarku, Florinda},
  title = {VOLUME COMPUTED TOMOGRAPHY DOSE INDEX EVALUATION FOR CT SCANNERS OPERATING IN VARIOUS RADIOLOGY DEPARTMENTS ACROSS ALBANIA},
  journal = {International Journal of Ecosystems and Ecology Science (IJEES)},
  year = {2026},
  volume = {16},
  number = {3},
  pages = {27--32},
  doi = {10.31407/ijees16.304}
}
Kostenlos starten →