Development and experimental validation of a low-cost CMOS based spectrometer for optical and fluorescence applications
Revista Mexicana de Física · 09.03.2026 · 72 (2 Mar-Apr)
DOI: 10.31349/revmexfis.72.020901 ↗Kurz gefasst
Fachartikel in der Zeitschrift Revista Mexicana de Física, online erschienen am 09.03.2026. Verfasst von Carlos Alberto Hernández Gutiérrez und 1 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
This work presents the design, construction, and experimental validation of a low-cost CMOS-based spectrometer built using a 3D-printed structure, a general-purpose webcam, and a low-cost diffraction grating. The system was calibrated using LED light sources and validated against a commercial Ocean HR4000CG-UV-NIR spectrometer. Results show that the CMOS-based spectrometer achieves wavelength detection accuracy within ±7 nm and is capable of resolving emission spectra with ~1 nm resolution spacing in the visible range (440–620 nm). The device was also applied to measure fluorescence in vitamin-enriched beverages and to analyze the absorbance properties of colored filters, demonstrating its utility in both chemical and optoelectronic applications. These findings confirm the feasibility of implementing low-cost spectroscopic tools for educational and scientific use.
Zitieren
APA
Gutiérrez, C. A. H., & Araujo, J. M. (2026). Development and experimental validation of a low-cost CMOS based spectrometer for optical and fluorescence applications. Revista Mexicana de Física, 72(2 Mar-Apr). https://doi.org/10.31349/revmexfis.72.020901
Vancouver
Gutiérrez CAH, Araujo JM. Development and experimental validation of a low-cost CMOS based spectrometer for optical and fluorescence applications. Revista Mexicana de Física. 2026;72(2 Mar-Apr). doi:10.31349/revmexfis.72.020901
BibTeX
@article{Gutirrez2026,
author = {Gutiérrez, Carlos Alberto Hernández and Araujo, J-E Moreno},
title = {Development and experimental validation of a low-cost CMOS based spectrometer for optical and fluorescence applications},
journal = {Revista Mexicana de Física},
year = {2026},
volume = {72},
number = {2 Mar-Apr},
doi = {10.31349/revmexfis.72.020901}
}
EUVIMED