A Comparative Study of Automated Blood Pressure Device and an Approved Standard Blood Pressure Measuring Device in Young Healthy Population
Journal of Nepalgunj Medical College · 31.12.2025 · 23 (2) · 6-9
DOI: 10.3126/jngmc.v23i2.90537 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nepalgunj Medical College, online erschienen am 31.12.2025. Verfasst von Rakesh Kumar Jha und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).
Abstract (Originalsprache)
Introduction
Blood pressure is a key indicator of cardiovascular health. Automated devices are widely used for convenience, but their accuracy compared with the auscultatory method in young healthy adults remains under investigation.
Aims
To determine whether automated and auscultatory blood pressure measurement methods show clinically meaningful differences in a young healthy cohort and whether sex or arm differences influence readings.
Methods
A comparative cross-sectional study using purposive sampling was conducted among 80 medical students aged 20–30 years. Although student samples may not represent the general population, they provide controlled and stable baseline physiological conditions for comparative device evaluation. Blood pressure was measured three times per arm using an automated device and a standard auscultatory method under stable room temperature, controlled lighting, and a quiet environment.
Results
Systolic blood pressure measured by the auscultatory method was significantly higher than that obtained by the automated device (right arm: mean difference 2.94 mmHg; left arm: 2.50 mmHg). Diastolic blood pressure differences were not statistically significant. Inter-arm differences were minimal. Males had higher systolic BP than females across both arms, while diastolic differences were negligible.
Conclusion
Automated devices slightly underestimate systolic blood pressure compared with the auscultatory method in young healthy adults, while diastolic readings are comparable. Automated monitors are suitable for screening and home monitoring, but clinical decisions near systolic thresholds should preferentially rely on validated auscultatory measurements.
Zitieren
APA
Jha, R. K., Shrewastwa, M. K., & Poudel, A. (2025). A Comparative Study of Automated Blood Pressure Device and an Approved Standard Blood Pressure Measuring Device in Young Healthy Population. Journal of Nepalgunj Medical College, 23(2), 6-9. https://doi.org/10.3126/jngmc.v23i2.90537
Vancouver
Jha RK, Shrewastwa MK, Poudel A. A Comparative Study of Automated Blood Pressure Device and an Approved Standard Blood Pressure Measuring Device in Young Healthy Population. Journal of Nepalgunj Medical College. 2025;23(2):6-9. doi:10.3126/jngmc.v23i2.90537
BibTeX
@article{Jha2025,
author = {Jha, Rakesh Kumar and Shrewastwa, Mukesh Kumar and Poudel, Abhishek},
title = {A Comparative Study of Automated Blood Pressure Device and an Approved Standard Blood Pressure Measuring Device in Young Healthy Population},
journal = {Journal of Nepalgunj Medical College},
year = {2025},
volume = {23},
number = {2},
pages = {6--9},
doi = {10.3126/jngmc.v23i2.90537}
}
EUVIMED