Fluorescence Immunoassay versus HPLC: Two Approaches for HbA1c Measurement, Same Reliability?
GSC Biological and Pharmaceutical Sciences · 31.05.2026 · 35 (2) · 316-321
DOI: 10.30574/gscbps.2026.35.2.0201 ↗Kurz gefasst
Fachartikel in der Zeitschrift GSC Biological and Pharmaceutical Sciences, online erschienen am 31.05.2026. Verfasst von Mansar Nouha und 3 weiteren Autor:innen.
Abstract (Originalsprache)
Objective
To compare the analytical performance of two HbA1c measurement methods: high-performance liquid chromatography (HPLC) using the Turbo Variant II system (Bio-Rad), considered the reference method, and fluorescence immunoassay using the STANDARD F200 system (SD Biosensor).
Methods
Fifty-five EDTA-anticoagulated whole blood samples were analyzed using both techniques. Results were compared using Bland-Altman analysis and Passing-Bablok regression to assess agreement and correlation between the two methods.
Results
The mean HbA1c values were 7.461 ± 2.037% for HPLC and 7.740 ± 2.021% for the STANDARD F200. Bland-Altman analysis showed a mean bias of 0.279%, with limits of agreement ranging from -0.248% to 0.804%. Passing-Bablok regression yielded the equation Y = 0.9837X + 0.3999 with a high correlation coefficient (r = 0.991, p < 0.001).
Conclusion
Both methods demonstrated good correlation and satisfactory agreement, with a slight overestimation of values by the STANDARD F200. Given its faster processing time and greater accessibility, the STANDARD F200 represents a reliable alternative to HPLC, particularly in settings requiring rapid screening or with limited resources.
Zitieren
APA
Nouha, M., Insaf, H., Abderrahman, B., & Saliha, C. (2026). Fluorescence Immunoassay versus HPLC: Two Approaches for HbA1c Measurement, Same Reliability?. GSC Biological and Pharmaceutical Sciences, 35(2), 316-321. https://doi.org/10.30574/gscbps.2026.35.2.0201
Vancouver
Nouha M, Insaf H, Abderrahman B, Saliha C. Fluorescence Immunoassay versus HPLC: Two Approaches for HbA1c Measurement, Same Reliability?. GSC Biological and Pharmaceutical Sciences. 2026;35(2):316-321. doi:10.30574/gscbps.2026.35.2.0201
BibTeX
@article{Nouha2026,
author = {Nouha, Mansar and Insaf, Haddi and Abderrahman, Boukhira and Saliha, Chellak},
title = {Fluorescence Immunoassay versus HPLC: Two Approaches for HbA1c Measurement, Same Reliability?},
journal = {GSC Biological and Pharmaceutical Sciences},
year = {2026},
volume = {35},
number = {2},
pages = {316--321},
doi = {10.30574/gscbps.2026.35.2.0201}
}
EUVIMED