Automated Total-Body Perfusion Imaging with 15 O-Water PET Using Basis Functions and Organ-Specific Model Selection
Journal of Nuclear Medicine · 2025
DOI: 10.2967/jnumed.124.269409 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine, 2025. Verfasst von Anna Åhlström und 7 weiteren Autor:innen. Beteiligte Einrichtung: Molecular Imaging and Medical Physics, Department of Surgical Sciences, Uppsala University, Uppsala, Sweden; and anna.a.ahlstrom@uu.se. Hauptthemen laut MEDLINE: Oxygen Radioisotopes, Whole Body Imaging, Water, Perfusion Imaging. Die Arbeit wurde bisher 4-mal zitiert (Crossref) und nennt 45 Literaturangaben.
Abstract (Originalsprache)
Long-axial-field-of-view PET with 15O-water allows perfusion to be measured in the whole body simultaneously. The purpose of this work was to describe a method for automated computation of total-body parametric perfusion images using PET information only and to validate the perfusion and volume of distribution (V T) values obtained by comparing them with the gold standard (nonlinear regression analysis).
Methods
Data from 10 subjects at Turku PET Centre were evaluated. Each subject underwent a 4-min 40-s dynamic PET/CT scan starting simultaneously with a controlled bolus administration of 350 MBq of 15O-water. Arterial and venous blood curves were defined using cluster analysis. Delay correction was performed by down-sampling the PET volume, using nonlinear regression for estimation of the delay for each subvolume, interpolation of delay values to the original matrix, and delay correction of all voxel time-activity curves, allowing for linearization of the model. Total-body perfusion images were calculated using several basis function implementations of the single-tissue-compartment model, considering the variations in blood supply to different organs. Model selection for each voxel was performed using cluster analysis to identify different organs. Perfusion and V T values based on the automated parametric imaging method were validated by comparison of mean organ values with nonlinear regression of the appropriate compartment models to whole-organ time-activity curves.
Results
The results showed good agreement between the parameters achieved from the automated parametric images and nonlinear regression. Correlation (R 2) and agreement between linear and nonlinear analyses were high, with an R 2 of 0.99 for both perfusion and V T, with a slope of 0.98 and 1.01 for perfusion and V T, respectively.
Conclusion
Perfusion and V T values based on automated total-body parametric analysis agreed well with those based on nonlinear regression of whole-organ time-activity curves.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Autor:innen und Institutionen
- Molecular Imaging and Medical Physics, Department of Surgical Sciences, Uppsala University, Uppsala, Sweden; and anna.a.ahlstrom@uu.se.
- Molecular Imaging and Medical Physics, Department of Surgical Sciences, Uppsala University, Uppsala, Sweden; and.
- Turku PET Centre, University of Turku and Turku University Hospital, Turku, Finland.
Zitieren
APA
Åhlström, A., Lindström, E., Maaniitty, T., Iida, H., Kärpijoki, H., Sörensen, J., Knuuti, J., & Lubberink, M. (2025). Automated Total-Body Perfusion Imaging with 15 O-Water PET Using Basis Functions and Organ-Specific Model Selection. Journal of Nuclear Medicine. https://doi.org/10.2967/jnumed.124.269409
Vancouver
Åhlström A, Lindström E, Maaniitty T, Iida H, Kärpijoki H, Sörensen J, et al. Automated Total-Body Perfusion Imaging with 15 O-Water PET Using Basis Functions and Organ-Specific Model Selection. Journal of Nuclear Medicine. 2025. doi:10.2967/jnumed.124.269409
BibTeX
@article{hlstrm2025,
author = {Åhlström, Anna and Lindström, Elin and Maaniitty, Teemu and Iida, Hidehiro and Kärpijoki, Henri and Sörensen, Jens and Knuuti, Juhani and Lubberink, Mark},
title = {Automated Total-Body Perfusion Imaging with 15 O-Water PET Using Basis Functions and Organ-Specific Model Selection},
journal = {Journal of Nuclear Medicine},
year = {2025},
doi = {10.2967/jnumed.124.269409}
}
EUVIMED