Automated Quantitative Analysis of American College of Radiology PET Phantom Images
Journal of Nuclear Medicine Technology · 2019-09 · 47 (3) · 249-254
DOI: 10.2967/jnmt.118.221317 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Nuclear Medicine Technology, online erschienen am 24.04.2019. Verfasst von Frank P DiFilippo und 2 weiteren Autor:innen. Beteiligte Einrichtung: Department of Nuclear Medicine, Cleveland Clinic, Cleveland, Ohio difilif@ccf.org. Hauptthemen laut MEDLINE: Phantoms, Imaging, Societies, Medical. Die Arbeit wurde bisher 9-mal zitiert (Crossref) und nennt 21 Literaturangaben.
Abstract (Originalsprache)
Evaluation of PET image quality is central to annual physics surveys, quality assurance, and laboratory accreditation. A common method is to image the American College of Radiology (ACR) PET phantom, which contains hot and cold structures of various sizes in a warm background. Performance evaluation involves qualitative assessment of hot and cold structure visibility and overall image quality. Some criteria are quantitative and rely on manually drawn regions of interest (ROIs) to measure SUV. Fully automated scoring of ACR PET phantom images would improve efficiency, avoid observer-related dependencies, and possibly provide more robust evaluation of image quality.
Methods
Software was developed to coregister PET images to a phantom template and to compute ROI measurements of hot vial activity (SUVmax) and background activity (SUVmean) automatically. In addition, 3-dimensional volumes of interest (VOIs) were generated to measure hot vial activity (SUVvial), background activity, and cold rod contrast. Consistency of the ROI-based and VOI-based methods was evaluated using phantom data from a total of 17 annual physics surveys of 3 PET/CT scanners with the same PET detector design.
Results
The automated software processed all PET phantom datasets successfully. SUV consistency for hot vials was improved through use of cylindric VOIs and through normalization with respect to assayed activities and dilution volumes used in phantom preparation. Average vial SUV SD improved from 8.0% for standard SUVmax to 3.2% for normalized SUVvial Similarly, the SD for the SUV ratio of 16- to 25-mm vials improved from 5.0% for SUVmax to 3.2% for SUVvial Background SUVmean had a similar consistency between the ROI and VOI methods. Cold rod contrast was highly consistent, offering a potential alternative to qualitative visual assessment of low-contrast performance.
Conclusion
Automated quantitative scoring of the ACR PET phantom is feasible and offers the advantages of more efficient, consistent, and thorough performance characterization. Acceptance ranges for SUVs and ratios likely can be tightened if normalized VOI measurements are used. Further testing with phantom data from a variety of PET scanners is necessary to establish suitable quantitative thresholds for acceptable performance.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Nuclear Medicine, Cleveland Clinic, Cleveland, Ohio difilif@ccf.org.
- Department of Nuclear Medicine, Cleveland Clinic, Cleveland, Ohio.
Zitieren
APA
DiFilippo, F. P., Patel, M., & Patel, S. (2019). Automated Quantitative Analysis of American College of Radiology PET Phantom Images. Journal of Nuclear Medicine Technology, 47(3), 249-254. https://doi.org/10.2967/jnmt.118.221317
Vancouver
DiFilippo FP, Patel M, Patel S. Automated Quantitative Analysis of American College of Radiology PET Phantom Images. Journal of Nuclear Medicine Technology. 2019;47(3):249-254. doi:10.2967/jnmt.118.221317
BibTeX
@article{DiFilippo2019,
author = {DiFilippo, Frank P and Patel, Meghal and Patel, Sagar},
title = {Automated Quantitative Analysis of American College of Radiology PET Phantom Images},
journal = {Journal of Nuclear Medicine Technology},
year = {2019},
volume = {47},
number = {3},
pages = {249--254},
doi = {10.2967/jnmt.118.221317}
}
EUVIMED