Automated quality control analysis for American College of Radiology (ACR) digital mammography (DM) phantom images
Journal of Applied Clinical Medical Physics · 2024
DOI: 10.1002/acm2.14548 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Applied Clinical Medical Physics, 2024. Verfasst von Zeyad Alawaji und 3 weiteren Autor:innen. Beteiligte Einrichtung: Discipline of Medical Imaging Science, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia. Hauptthemen laut MEDLINE: Mammography, Phantoms, Imaging, Quality Control, Image Processing, Computer-Assisted. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 30 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Purpose
To develop and validate an automated software analysis method for mammography image quality assessment of the American College of Radiology (ACR) digital mammography (DM) phantom images.
Methods
Twenty‐seven DICOM images were acquired using Fuji mammography systems. All images were evaluated by three expert medical physicists using the Royal Australian and New Zealand College of Radiologists (RANZCR) mammography quality control guideline. To enhance the robustness and sensitivity assessment of our algorithm, an additional set of 12 images from a Hologic mammography system was included to test various phantom positional adjustments. The software automatically chose multiple regions of interest (ROIs) for analysis. A template matching method was primarily used for image analysis, followed by an additional method that locates and scores each target object (speck groups, fibers, and masses).
Results
The software performance shows a good to excellent agreement with the average scoring of observers (intraclass correlation coefficient [ICC] of 0.75, 0.79, 0.82 for speck groups, fibers, and masses, respectively). No significant differences were found in the scoring of target objects between human observers and the software. Both methods achieved scores meeting the pass criteria for speck groups and masses. Expert observers allocated lower scores to fiber objects, with diameters less than 0.61 mm, when compared to the software. The software was able to accurately score objects when the phantom position changed by up to 25 mm laterally, up to 5 degrees rotation, and overhanging the chest wall edge by up to 15 mm.
Conclusions
Automated software analysis is a feasible method that may help improve the consistency and reproducibility of mammography image quality assessment with reduced reliance on human interaction and processing time.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Discipline of Medical Imaging Science, Faculty of Medicine and Health, The University of Sydney, Sydney, New South Wales, Australia.
- Department of Radiologic Technology, College of Applied Medical Sciences, Qassim University, Buraydah, Saudi Arabia.
- Medical Physics Department, Western Sydney Local Health District, Westmead, New South Wales, Australia.
- Medical Imaging Department, Prince of Wales Hospital, Randwick, New South Wales, Australia.
ORCID-Profile: Zeyad Alawaji
Zitieren
APA
Alawaji, Z., Taba, S. T., Cartwright, L., & Rae, W. (2024). Automated quality control analysis for American College of Radiology (ACR) digital mammography (DM) phantom images. Journal of Applied Clinical Medical Physics. https://doi.org/10.1002/acm2.14548
Vancouver
Alawaji Z, Taba ST, Cartwright L, Rae W. Automated quality control analysis for American College of Radiology (ACR) digital mammography (DM) phantom images. Journal of Applied Clinical Medical Physics. 2024. doi:10.1002/acm2.14548
BibTeX
@article{Alawaji2024,
author = {Alawaji, Zeyad and Taba, Seyedamir Tavakoli and Cartwright, Lucy and Rae, William},
title = {Automated quality control analysis for American College of Radiology (ACR) digital mammography (DM) phantom images},
journal = {Journal of Applied Clinical Medical Physics},
year = {2024},
doi = {10.1002/acm2.14548}
}
EUVIMED