Validation of the Omron HBP-9031C blood pressure monitor for clinics and hospitals according to the ANSI/AAMI/ISO 81060-2:2013 protocol

Kanako Saito, Yukiko Hishiki, Hakuo Takahashi

Annals of Clinical Hypertension · 2019

DOI: 10.29328/journal.ach.1001018 ↗

Kurz gefasst

Studienprotokoll in der Zeitschrift Annals of Clinical Hypertension, 2019. Verfasst von Kanako Saito und 2 weiteren Autor:innen. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 7 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Objective

The present study aimed to evaluate the accuracy of the Omron HBP-9031C automated oscillometric upper-arm blood pressure (BP) measurement device for blood pressure monitoring, according to the ANSI/AAMI/ISO 81060-2:2013 protocol (ANSI/AAMI/ISO). Participants and Method: The device evaluations were performed in 85 participants, who fulfilled the inclusion criteria of the protocol. The validation procedure and data analysis followed the protocol precisely.

Results

In the ANSI/AAMI/ISO 81060-2-2013 validation procedure (criterion 1), the mean ± SD of the differences between the test device and reference BP was 0.5 ± 7.84/-1.9 ± 6.30 mmHg (systolic/diastolic). The mean differences between the two observers and the Omron HBP-9031C were 0.5 ± 6.69 mmHg (range, −18 to 15 mmHg) for systolic BP and -1.9 ± 5.63 mmHg (range, −17 to 14 mmHg) for diastolic BP, according to criterion 2. The two criteria of the ANSI/AAMI/ISO were fulfilled.

Conclusion

The professional OMRON BP monitor, HBP-9031C fulfilled the requirements of the ANSI/AAMI/ISO validation standard and can be recommended for clinical use.

Zitieren

APA

Saito, K., Hishiki, Y., & Takahashi, H. (2019). Validation of the Omron HBP-9031C blood pressure monitor for clinics and hospitals according to the ANSI/AAMI/ISO 81060-2:2013 protocol. Annals of Clinical Hypertension. https://doi.org/10.29328/journal.ach.1001018

Vancouver

Saito K, Hishiki Y, Takahashi H. Validation of the Omron HBP-9031C blood pressure monitor for clinics and hospitals according to the ANSI/AAMI/ISO 81060-2:2013 protocol. Annals of Clinical Hypertension. 2019. doi:10.29328/journal.ach.1001018

BibTeX
@article{Saito2019,
  author = {Saito, Kanako and Hishiki, Yukiko and Takahashi, Hakuo},
  title = {Validation of the Omron HBP-9031C blood pressure monitor for clinics and hospitals according to the ANSI/AAMI/ISO 81060-2:2013 protocol},
  journal = {Annals of Clinical Hypertension},
  year = {2019},
  doi = {10.29328/journal.ach.1001018}
}
Kostenlos starten →