Accuracy Evaluation of the Omron HEM‐7120 Automated Oscillometric Upper‐arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060–2:2018 Numerical Criteria

Tipanan Pisithkul iD, Kwanchai Pirojsakul iD

The Journal of Clinical Hypertension · 2026-10 · 28 (10)

DOI: 10.1111/jch.70370 ↗

Kurz gefasst

Fachartikel in der Zeitschrift The Journal of Clinical Hypertension, online erschienen am 27.10.2026. Verfasst von Tipanan Pisithkul und 1 weiteren Autor:innen. Beteiligte Einrichtung: Division of Nephrology, Department of Pediatrics, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok, Thailand. Hauptthemen laut MEDLINE: Oscillometry, Blood Pressure Monitors, Blood Pressure Monitoring, Ambulatory, Hypertension. Die Arbeit nennt 18 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC-ND 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

This study evaluated the Omron HEM‐7120 automated oscillometric upper‐arm home blood pressure monitor in children and adolescents against the numerical accuracy criteria of the AAMI/ESH/ISO Universal Standard. Test readings were compared with mercury auscultatory readings from two physicians using the same‐arm sequential method. Three cuffs covered arm circumferences of 12–18, 17–22, and 22–32 cm. Of 90 participants, 85 were analyzed, yielding 255 comparisons. Mean differences were −4.0 ± 5.94 mmHg for systolic and −1.8 ± 7.20 mmHg for diastolic blood pressure, meeting Criterion 1. Participant standard deviations were 5.07 and 6.48 mmHg, below Criterion 2 limits. However, the requirements regarding age, reference‐blood‐pressure and cuff distributions, reference‐measurement variability, and observer qualification were not fully met. The device met numerical criteria in this cohort, but it is not fully validated under the Universal Standard.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

OscillometryBlood Pressure MonitorsBlood Pressure Monitoring, AmbulatoryHypertensionHumansChildAdolescentFemaleMaleBlood PressureReproducibility of ResultsBlood Pressure DeterminationArm

Stichwörter der Autor:innen

AAMI/ESH/ISOOscillometric blood pressure monitoradolescentsblood pressure monitoringchildrendevice validationhome blood pressure monitoring

Autor:innen und Institutionen

  • Division of Nephrology, Department of Pediatrics, Faculty of Medicine Ramathibodi Hospital, Mahidol University, Bangkok, Thailand.

ORCID-Profile: Tipanan Pisithkul, Kwanchai Pirojsakul

Zitieren

APA

Pisithkul, T., & Pirojsakul, K. (2026). Accuracy Evaluation of the Omron HEM‐7120 Automated Oscillometric Upper‐arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060–2:2018 Numerical Criteria. The Journal of Clinical Hypertension, 28(10). https://doi.org/10.1111/jch.70370

Vancouver

Pisithkul T, Pirojsakul K. Accuracy Evaluation of the Omron HEM‐7120 Automated Oscillometric Upper‐arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060–2:2018 Numerical Criteria. The Journal of Clinical Hypertension. 2026;28(10). doi:10.1111/jch.70370

BibTeX
@article{Pisithkul2026,
  author = {Pisithkul, Tipanan and Pirojsakul, Kwanchai},
  title = {Accuracy Evaluation of the Omron HEM‐7120 Automated Oscillometric Upper‐arm Home Blood Pressure Monitor in Children and Adolescents Using the AAMI/ESH/ISO 81060–2:2018 Numerical Criteria},
  journal = {The Journal of Clinical Hypertension},
  year = {2026},
  volume = {28},
  number = {10},
  doi = {10.1111/jch.70370}
}
Kostenlos starten →