Enhancing IV Smart Pump Setup and Flow Rate Accuracy: An Exploratory Study on an Educational Intervention in Critical Care
International Journal of Critical Care · 12.05.2026 · 20 (2) · 37-46
DOI: 10.29173/ijcc1050 ↗Kurz gefasst
Fachartikel in der Zeitschrift International Journal of Critical Care, online erschienen am 12.05.2026. Verfasst von Sarah Romain und 2 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).
Abstract (Originalsprache)
Background
Large-volume intravenous (IV) smart pumps (IVSP) are crucial for IV medication administration in acute and critical care. The Dose Error Reduction System (DERS) helps prevent dosing errors but does not address actual versus programmed flow rate discrepancies, which are influenced by various clinical and technical factors.
Aim
This study aimed to (1) observe adherence to manufacturer setup requirements and IV flow rate accuracy, (2) implement an educational program for critical care nurses, and (3) reassess adherence and accuracy post-education.
Methods
Observations were collected in critically ill patients in two phases: 25 before and 25 after the educational program. The program was delivered during staff meetings and via recorded sessions, focusing on manufacturer-required setup for head-height dependent IVSP. Results/Findings: Key findings included improved compliance with unclamped secondary tubing (72.7% pre-education vs. 91.7% post-education) and increased use of back-check valves (93.1% vs. 99.0%). Median absolute flow rate deviations decreased from 2.0 mL/hr to 1.3 mL/hr for primary infusions and from 6.0 mL/hr to 4.0 mL/hr for secondary infusions.
Conclusions
The educational program improved certain setup practices, but overall compliance remained suboptimal. Future research should include a larger sample size and consider additional clinical factors affecting IV flow rate accuracy.
Zitieren
APA
Romain, S., Vital, C., & Giuliano, K. (2026). Enhancing IV Smart Pump Setup and Flow Rate Accuracy: An Exploratory Study on an Educational Intervention in Critical Care. International Journal of Critical Care, 20(2), 37-46. https://doi.org/10.29173/ijcc1050
Vancouver
Romain S, Vital C, Giuliano K. Enhancing IV Smart Pump Setup and Flow Rate Accuracy: An Exploratory Study on an Educational Intervention in Critical Care. International Journal of Critical Care. 2026;20(2):37-46. doi:10.29173/ijcc1050
BibTeX
@article{Romain2026,
author = {Romain, Sarah and Vital, Cidalia and Giuliano, Karen},
title = {Enhancing IV Smart Pump Setup and Flow Rate Accuracy: An Exploratory Study on an Educational Intervention in Critical Care},
journal = {International Journal of Critical Care},
year = {2026},
volume = {20},
number = {2},
pages = {37--46},
doi = {10.29173/ijcc1050}
}
EUVIMED