EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Data-based program management of system-wide IV smart pump integration

Karen K Giuliano, Rebecca S Mahuren, Jacob Balyeat

American Journal of Health-System Pharmacy · 01.01.2024 · 81 (1) · e30-e36

DOI: 10.1093/ajhp/zxad245 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Health-System Pharmacy, online erschienen am 07.10.2023. Verfasst von Karen K Giuliano und 2 weiteren Autor:innen. Beteiligte Einrichtung: Elaine Marieb Center for Nursing and Engineering Innovation, University of Massachusetts Amherst, Amherst, MA, USA. Hauptthemen laut MEDLINE: Medication Errors, Patient Safety. Die Arbeit wurde bisher 22-mal zitiert (Crossref) und nennt 20 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Purpose

Smart pump bidirectional interoperability offers automated infusion programming and documentation that can improve patient safety and workflow efficiency. This technology has been poorly implemented across US hospitals, and there is little guidance on the tracking or monitoring of interoperability systems. The purpose of this report is to describe the successful implementation of intravenous (IV) smart pump interoperability in a large health system. Summary Bidirectional IV smart pump interoperability and compliance monitoring were implemented across a large Midwestern health system using ICU Medical’s Plum 360 and LifeCare PCA devices and Smith Medical’s MedFusion 4000 Syringe Pump devices. The hospital system’s experience in implementing and monitoring IV smart pump compliance using automated reports and a dedicated medication safety integration nurse is described. Compliance trends suggest that the implementation of IV smart pump interoperability has achieved a reduction in programming outside of the dose error reduction system, manual overrides, and IV medication administration error rates.

Conclusion

The monitoring of smart pump compliance has had demonstrated benefits in investigating usability concerns, recognizing system errors, and identifying increased needs for nurse training. This program can serve as an example for other healthcare systems adopting IV smart pump interoperability.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Medication ErrorsPatient SafetyHumansAdministration, IntravenousInfusion PumpsHospitalsInfusions, Intravenous

Stichwörter der Autor:innen

auto-documentationauto-programminghealth information interoperabilityinfusion pumpsmedication errorpatient safety

Autor:innen und Institutionen

  • Elaine Marieb Center for Nursing and Engineering Innovation, University of Massachusetts Amherst, Amherst, MA, USA.
  • Parkview Health, Coldwater, MI, USA.
  • Parkview Regional Medical Center, Fort Wayne, IN, USA.

Zitieren

APA

Giuliano, K. K., Mahuren, R. S., & Balyeat, J. (2024). Data-based program management of system-wide IV smart pump integration. American Journal of Health-System Pharmacy, 81(1), e30-e36. https://doi.org/10.1093/ajhp/zxad245

Vancouver

Giuliano KK, Mahuren RS, Balyeat J. Data-based program management of system-wide IV smart pump integration. American Journal of Health-System Pharmacy. 2024;81(1):e30-e36. doi:10.1093/ajhp/zxad245

BibTeX
@article{Giuliano2024,
  author = {Giuliano, Karen K and Mahuren, Rebecca S and Balyeat, Jacob},
  title = {Data-based program management of system-wide IV smart pump integration},
  journal = {American Journal of Health-System Pharmacy},
  year = {2024},
  volume = {81},
  number = {1},
  pages = {e30--e36},
  doi = {10.1093/ajhp/zxad245}
}
Kostenlos starten →