The right leg drive (RLD) is a circuit associated with electrocardiography acquisition circuits. For electromyography (EMG), the RLD circuit is used to a lesser degree. In general, the RLD circuit provides better noise reduction. This study compares the output of the EMG with and without the RLD circuit. The results indicate that with a good filter design, the direct grounding method can match the RLD in terms of noise reduction. As a result, EMG application, the RLD drive can be omitted.