EFFECT of VARIATION of C-ARM ANGLE POSITION to DOSE RATE was RECEIVED in SURGICAL PROCEDURE at the CENTRAL SURGICAL INSTALLATION

Isti Nur Subkhiyati, Wahyu Setia Budi, Heri Sutanto

International Journal of Innovative Research in Advanced Engineering · 2017

DOI: 10.26562/ijirae.2017.jyae10083 ↗

Kurz gefasst

Fachartikel in der Zeitschrift International Journal of Innovative Research in Advanced Engineering, 2017. Verfasst von Isti Nur Subkhiyati und 2 weiteren Autor:innen.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

An anthropomorphic phantom dose rate measurement has been done in C-Arm room at Central Surgical Installation with surveymeter. Measurements were made with 8 variations of the C-Arm fluoroscopy angle commonly used in the operating procedure, as well as the 1 meter point of measurement against the radiation source. The variations of these angles are 00, 450, 900, 1350, 1800, 2250, 2700, and 3150. The dose rate measurements at a distance of 1 m from the radiation source with angle variations are 380 μSv/h, 430 μSv/h, 680 μSv/h, 29 μSv/h, 220 μSv/h, 350 μSv/h, 1370 μSv/h and 1020 μSv/h. The measurement results showed that the highest dose rate at the C-arm angle of 2700 and the lowest dose at the C-Arm 1350 angle position. From the measurement results can be seen the effect of angular position used with the acceptable dose rate, the lowest to highest dose rate received based on the angular position in sequence are 1350, 1800, 2250, 00, 450, 900, 3150, 2700. Keywords— C-Arm, Laju dosis, Fluoroskopi, Phantom Antropomorfik, Titik pengukuran

Zitieren

APA

Subkhiyati, I. N., Budi, W. S., & Sutanto, H. (2017). EFFECT of VARIATION of C-ARM ANGLE POSITION to DOSE RATE was RECEIVED in SURGICAL PROCEDURE at the CENTRAL SURGICAL INSTALLATION. International Journal of Innovative Research in Advanced Engineering. https://doi.org/10.26562/ijirae.2017.jyae10083

Vancouver

Subkhiyati IN, Budi WS, Sutanto H. EFFECT of VARIATION of C-ARM ANGLE POSITION to DOSE RATE was RECEIVED in SURGICAL PROCEDURE at the CENTRAL SURGICAL INSTALLATION. International Journal of Innovative Research in Advanced Engineering. 2017. doi:10.26562/ijirae.2017.jyae10083

BibTeX
@article{Subkhiyati2017,
  author = {Subkhiyati, Isti Nur and Budi, Wahyu Setia and Sutanto, Heri},
  title = {EFFECT of VARIATION of C-ARM ANGLE POSITION to DOSE RATE was RECEIVED in SURGICAL PROCEDURE at the CENTRAL SURGICAL INSTALLATION},
  journal = {International Journal of Innovative Research in Advanced Engineering},
  year = {2017},
  doi = {10.26562/ijirae.2017.jyae10083}
}
Kostenlos starten →