Assessing the validity of inertial measurement units for shoulder kinematics using a commercial sensor-software system

Jakob Henschke, Hannes Kaplick, Monique Wochatz, Tilman Engel

publish.UP (University of Potsdam) · 01.01.2023

DOI: 10.25932/publishup-57827 ↗

Kurz gefasst

Fachartikel in der Zeitschrift publish.UP (University of Potsdam), erschienen am 01.01.2023. Verfasst von Jakob Henschke und 3 weiteren Autor:innen.

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background and Aims Wearable inertial sensors may offer additional kinematic parameters of the shoulder compared to traditional instruments such as goniometers when elaborate and time-consuming data processing procedures are undertaken. However, in clinical practice simple-real time motion analysis is required to improve clinical reasoning. Therefore, the aim was to assess the criterion validity between a portable "off-the-shelf" sensor-software system (IMU) and optical motion (Mocap) for measuring kinematic parameters during active shoulder movements.

Methods

24 healthy participants (9 female, 15 male, age 29 +/- 4 years, height 177 +/- 11 cm, weight 73 +/- 14 kg) were included. Range of motion (ROM), total range of motion (TROM), peak and mean angular velocity of both systems were assessed during simple (abduction/adduction, horizontal flexion/horizontal extension, vertical flexion/extension, and external/internal rotation) and complex shoulder movements. Criterion validity was determined using intraclass-correlation coefficients (ICC), root mean square error (RMSE) and Bland and Altmann analysis (bias; upper and lower limits of agreement).

Results

ROM and TROM analysis revealed inconsistent validity during simple (ICC: 0.040-0.733, RMSE: 9.7 degrees-20.3 degrees, bias: 1.2 degrees-50.7 degrees) and insufficient agreement during complex shoulder movements (ICC: 0.104-0.453, RMSE: 10.1 degrees-23.3 degrees, bias: 1.0 degrees-55.9 degrees). Peak angular velocity (ICC: 0.202-0.865, RMSE: 14.6 degrees/s-26.7 degrees/s, bias: 10.2 degrees/s-29.9 degrees/s) and mean angular velocity (ICC: 0.019-0.786, RMSE:6.1 degrees/s-34.2 degrees/s, bias: 1.6 degrees/s-27.8 degrees/s) were inconsistent.

Conclusions

The "off-the-shelf" sensor-software system showed overall insufficient agreement with the gold standard. Further development of commercial IMU-software-solutions may increase measurement accuracy and permit their integration into everyday clinical practice.

Schlagwörter

inertial measurement unitshoulder kinematicscriterion validitywearable sensorsoptical motion capturerange of motionangular velocityshoulder movementclinical motion analysis

Zitieren

APA

Henschke, J., Kaplick, H., Wochatz, M., & Engel, T. (2023). Assessing the validity of inertial measurement units for shoulder kinematics using a commercial sensor-software system. publish.UP (University of Potsdam). https://doi.org/10.25932/publishup-57827

Vancouver

Henschke J, Kaplick H, Wochatz M, Engel T. Assessing the validity of inertial measurement units for shoulder kinematics using a commercial sensor-software system. publish.UP (University of Potsdam). 2023. doi:10.25932/publishup-57827

BibTeX
@article{Henschke2023,
  author = {Henschke, Jakob and Kaplick, Hannes and Wochatz, Monique and Engel, Tilman},
  title = {Assessing the validity of inertial measurement units for shoulder kinematics using a commercial sensor-software system},
  journal = {publish.UP (University of Potsdam)},
  year = {2023},
  doi = {10.25932/publishup-57827}
}
Kostenlos starten →