Multipurpose Radiofrequency Wire System Improves Procedural Workflow Compared With Mechanical Needle for Left Atrial Appendage Closure
Journal of Invasive Cardiology · 2025
DOI: 10.25270/jic/25.00179 ↗Kurz gefasst
Vergleichsstudie in der Zeitschrift Journal of Invasive Cardiology, 2025. Verfasst von Desiree Wussler und 4 weiteren Autor:innen. Beteiligte Einrichtung: Department of Cardiology, Vancouver General Hospital, University of British Columbia, Canada; Department of Cardiology and Cardiovascular Research In…. Hauptthemen laut MEDLINE: Atrial Appendage, Atrial Fibrillation, Workflow, Cardiac Catheterization.
Abstract (Originalsprache)
Percutaneous left atrial appendage closure (LAAC) is a rapidly emerging therapy for atrial fibrillation patients not suitable for anticoagulation. The study compared procedural time and adverse events between mechanical Brockenbrough needle and the VersaCross radiofrequency system for WATCHMAN implantation. All patients undergoing LAAC at Vancouver General Hospital were prospectively enrolled. Procedural workflow, times, and complications were compared in 20 patients: 10 using VersaCross radiofrequency system and 10 using Brockenbrough needle. Transseptal puncture success was 100%. The radiofrequency wire reduced total procedural time by 27% (21.5 vs 29.5 minutes; P=.008) and fluoroscopy and contrast use without increasing complications. One ischemic stroke occurred during 360-day follow-up unrelated to the device. The multipurpose radiofrequency system streamlines LAAC workflow and is a safe, efficient alternative to mechanical puncture. Larger randomized trials are warranted to confirm these benefits.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of Cardiology, Vancouver General Hospital, University of British Columbia, Canada; Department of Cardiology and Cardiovascular Research Institute Basel (CRIB), University Hospital Basel, University of Basel,…
- Department of Cardiology, Vancouver General Hospital, University of British Columbia, Canada.
- Department of Cardiology, Vancouver General Hospital, University of British Columbia, Canada. Email: jsaw@mail.ubc.ca.
Zitieren
APA
Wussler, D., Offen, S., Jelisejevas, J., Ogran, H., & Saw, J. (2025). Multipurpose Radiofrequency Wire System Improves Procedural Workflow Compared With Mechanical Needle for Left Atrial Appendage Closure. Journal of Invasive Cardiology. https://doi.org/10.25270/jic/25.00179
Vancouver
Wussler D, Offen S, Jelisejevas J, Ogran H, Saw J. Multipurpose Radiofrequency Wire System Improves Procedural Workflow Compared With Mechanical Needle for Left Atrial Appendage Closure. Journal of Invasive Cardiology. 2025. doi:10.25270/jic/25.00179
BibTeX
@article{Wussler2025,
author = {Wussler, Desiree and Offen, Sophie and Jelisejevas, Julius and Ogran, Hassan and Saw, Jacqueline},
title = {Multipurpose Radiofrequency Wire System Improves Procedural Workflow Compared With Mechanical Needle for Left Atrial Appendage Closure},
journal = {Journal of Invasive Cardiology},
year = {2025},
doi = {10.25270/jic/25.00179}
}