Evaluation of six sigma and Westgard rules: A comparative study of two hematology laboratories
Journal of Laboratory Physicians · 19.11.2026 · 1-7
DOI: 10.25259/jlp_196_2026 ↗Kurz gefasst
Fachartikel in der Zeitschrift Journal of Laboratory Physicians, online erschienen am 19.11.2026. Verfasst von Amelia Amel und 4 weiteren Autor:innen. Die Arbeit nennt 21 Literaturangaben.
Abstract (Originalsprache)
Objectives
The objectives of the study are to analyze internal quality-control results for mean corpuscular volume, mean corpuscular hemoglobin (MCH), and mean corpuscular hemoglobin concentration (MCHC) using Westgard rules and Six Sigma and to compare analytical performance between two type C hospital laboratories. Materials and Methods: This retrospective quantitative descriptive study used secondary daily quality-control data from Hospital X and Hospital Y from November 2025 to March 2026. All eligible low-, normal-, and high-level control results were included. Statistical analysis: The analysis included mean, standard deviation, bias, recovery percentage, coefficient of variation (CV), Levey-Jennings charts, Westgard rule events, and Six Sigma values.
Results
Most bias and recovery values met the pre-specified acceptance limits. The largest deviation occurred in the high-level MCHC control at Hospital Y, with a bias of 19.18% and recovery of 119.18%. Fifty-three Westgard rule events were identified: 52 1-2S warnings and one 2-2S rejection. Mean sigma was 6.47 at Hospital X and 4.96 at Hospital Y. MCH showed the best performance at Hospital X (sigma 8.53), whereas MCHC showed the lowest performance at Hospital Y (sigma 3.40).
Conclusions
Analytical performance differed by erythrocyte index and laboratory. Hospital X showed more stable analytical quality, particularly for MCH, whereas MCHC at Hospital Y required closer quality-control monitoring.
Zitieren
APA
Amel, A., Bastian, Nurhidayanti, Juraijin, D., & Verdiansyah (2026). Evaluation of six sigma and Westgard rules: A comparative study of two hematology laboratories. Journal of Laboratory Physicians, 0, 1-7. https://doi.org/10.25259/jlp_196_2026
Vancouver
Amel A, Bastian, Nurhidayanti, Juraijin D, Verdiansyah. Evaluation of six sigma and Westgard rules: A comparative study of two hematology laboratories. Journal of Laboratory Physicians. 2026;0:1-7. doi:10.25259/jlp_196_2026
BibTeX
@article{Amel2026,
author = {Amel, Amelia and Bastian and Nurhidayanti and Juraijin, Denny and Verdiansyah},
title = {Evaluation of six sigma and Westgard rules: A comparative study of two hematology laboratories},
journal = {Journal of Laboratory Physicians},
year = {2026},
volume = {0},
pages = {1--7},
doi = {10.25259/jlp_196_2026}
}
EUVIMED