Enhancing blood donor screening: A comparative study of portable hemoglobinometer and automated hematology analyzer

Bhargavi Gopaldas Rathod, Shivangi Ashokkumar Vaghela, Sangita Darshan Shah, Mamta C Shah, Nidhi M Bhatnagar, Shailesh P Kucha, Dhara J Shah, Priya G Kalavadiya

BJKines - National Journal of Basic & Applied Sciences · 17.06.2026 · 18 · 16-20

DOI: 10.25259/bjkines_15_2026 ↗

Kurz gefasst

Fachartikel in der Zeitschrift BJKines - National Journal of Basic & Applied Sciences, online erschienen am 17.06.2026. Verfasst von Bhargavi Gopaldas Rathod und 7 weiteren Autor:innen. Die Arbeit nennt 9 Literaturangaben.

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

Hemoglobin (Hb) estimation is a critical component of blood donor screening to ensure donor safety and adequate transfusion outcomes. Portable hemoglobinometers, such as HemoCue, offer rapid, point-of-care testing, while automated hematology analyzers are considered more accurate but less practical in field settings. To evaluate the accuracy, correlation, and practical applicability of hemoglobin estimation using a portable hemoglobinometer compared with an automated hematology analyzer. Material and Methods: A cross-sectional comparative study was conducted on 95 randomly selected blood donors. Capillary blood samples were analyzed using a portable hemoglobinometer (HemoCue), while venous blood samples were assessed using an automated hematology analyzer. Statistical analysis included mean, standard deviation, and 95% confidence intervals.

Results

The mean hemoglobin values obtained were 13.96 ± 1.81 g/dL (portable hemoglobinometer) and 14.14 ± 1.85 g/dL (automated hematology analyzer), with a mean difference of 0.2 g/dL. The results demonstrated a strong correlation and minimal variation between the two methods.

Conclusion

The portable hemoglobinometer provides reliable and accurate results comparable to automated analyzers, making it suitable for point-of-care donor screening, especially in resource-limited and field settings.

Zitieren

APA

Rathod, B. G., Vaghela, S. A., Shah, S. D., Shah, M. C., Bhatnagar, N. M., Kucha, S. P., Shah, D. J., & Kalavadiya, P. G. (2026). Enhancing blood donor screening: A comparative study of portable hemoglobinometer and automated hematology analyzer. BJKines - National Journal of Basic & Applied Sciences, 18, 16-20. https://doi.org/10.25259/bjkines_15_2026

Vancouver

Rathod BG, Vaghela SA, Shah SD, Shah MC, Bhatnagar NM, Kucha SP, et al. Enhancing blood donor screening: A comparative study of portable hemoglobinometer and automated hematology analyzer. BJKines - National Journal of Basic & Applied Sciences. 2026;18:16-20. doi:10.25259/bjkines_15_2026

BibTeX
@article{Rathod2026,
  author = {Rathod, Bhargavi Gopaldas and Vaghela, Shivangi Ashokkumar and Shah, Sangita Darshan and Shah, Mamta C and Bhatnagar, Nidhi M and Kucha, Shailesh P and Shah, Dhara J and Kalavadiya, Priya G},
  title = {Enhancing blood donor screening: A comparative study of portable hemoglobinometer and automated hematology analyzer},
  journal = {BJKines - National Journal of Basic & Applied Sciences},
  year = {2026},
  volume = {18},
  pages = {16--20},
  doi = {10.25259/bjkines_15_2026}
}
Kostenlos starten →