EUVIMEDEUVIMED entdecken
Open Access · CC BY 4.0

Technical Pitfalls in Optic Nerve Sheath Diameter POCUS: A Practical Checklist for Image Optimization

Ricardo Bañuelos-Huerta, Dibanhi S. Ruvalcaba-Trujillo, Jorge A. Lara-Arellano, Gilver Sebastián-Vázquez, Carlos Estrada-Valencia, Wendy J. Ramírez-Salvador, Daniel de J. Vázquez-Hernández, Montserrat Ramírez-Barranco, Eduardo Y. Aguilar-Lazcano, Wendy Sartillo-Guerra

POCUS Journal · 18.11.2026 · 11 (02) · 143-148

DOI: 10.24908/pocusj.v11i02.20719 ↗

Kurz gefasst

Fachartikel in der Zeitschrift POCUS Journal, online erschienen am 18.11.2026. Verfasst von Ricardo Bañuelos-Huerta und 9 weiteren Autor:innen. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Background

Point of care ultrasound (POCUS) measurement of the optic nerve sheath diameter (ONSD) has been established as a non-invasive method for estimating intracranial pressure in critically ill patients. However, the accuracy of the technique depends strongly on proper image acquisition. Errors in system optimization may result in inaccurate measurements and increased inter-operator variability.

Objective

Our objective was to describe the most common technical errors in ONSD measurement and to propose a checklist to optimize image acquisition.

Methods

A narrative review of the literature and an analysis of technical errors observed during POCUS training in critical care settings were completed. Emphasis was on ultrasound system parameters and image acquisition.

Results

Six key technical elements were identified to optimize ONSD measurement. They included preset selection, depth adjustment, focal zone positioning, transducer frequency, gain optimization, and identification of the central retinal artery and vein.

Conclusion

Standardization of ultrasound system settings may improve image quality and reduce measurement errors in ONSD, increasing reproducibility in emergency and critical care settings.

Zitieren

APA

Bañuelos-Huerta, R., Ruvalcaba-Trujillo, D. S., Lara-Arellano, J. A., Sebastián-Vázquez, G., Estrada-Valencia, C., Ramírez-Salvador, W. J., Vázquez-Hernández, D. D. J., Ramírez-Barranco, M., Aguilar-Lazcano, E. Y., & Sartillo-Guerra, W. (2026). Technical Pitfalls in Optic Nerve Sheath Diameter POCUS: A Practical Checklist for Image Optimization. POCUS Journal, 11(02), 143-148. https://doi.org/10.24908/pocusj.v11i02.20719

Vancouver

Bañuelos-Huerta R, Ruvalcaba-Trujillo DS, Lara-Arellano JA, Sebastián-Vázquez G, Estrada-Valencia C, Ramírez-Salvador WJ, et al. Technical Pitfalls in Optic Nerve Sheath Diameter POCUS: A Practical Checklist for Image Optimization. POCUS Journal. 2026;11(02):143-148. doi:10.24908/pocusj.v11i02.20719

BibTeX
@article{BauelosHuerta2026,
  author = {Bañuelos-Huerta, Ricardo and Ruvalcaba-Trujillo, Dibanhi S. and Lara-Arellano, Jorge A. and Sebastián-Vázquez, Gilver and Estrada-Valencia, Carlos and Ramírez-Salvador, Wendy J. and Vázquez-Hernández, Daniel de J. and Ramírez-Barranco, Montserrat and Aguilar-Lazcano, Eduardo Y. and Sartillo-Guerra, Wendy},
  title = {Technical Pitfalls in Optic Nerve Sheath Diameter POCUS: A Practical Checklist for Image Optimization},
  journal = {POCUS Journal},
  year = {2026},
  volume = {11},
  number = {02},
  pages = {143--148},
  doi = {10.24908/pocusj.v11i02.20719}
}
Kostenlos starten →