Intrauterine Pregnancy Detection and Gestational Age Assessment During Early Pregnancy by a Handheld Point-Of-Care Ultrasound Device Compared to a High-End Ultrasound System. An Accuracy and Reliability Study
POCUS Journal · 21.11.2022 · 7 (2) · 225-231
DOI: 10.24908/pocus.v7i2.15458 ↗Kurz gefasst
Fachartikel in der Zeitschrift POCUS Journal, online erschienen am 21.11.2022. Verfasst von Mariela Skendi und 6 weiteren Autor:innen. Beteiligte Einrichtung: Department of General Practice, Faculty of Medicine, Sorbonne University Paris France. Die Arbeit wurde bisher 8-mal zitiert (Crossref). Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Objective
The main objective of this study is the evaluation of the accuracy and reliability of a handheld point of care ultrasound device (POCUS-hd) for intrauterine pregnancy (IUP) detection compared to comprehensive reference transabdominal ultrasound (TU). The secondary objectives were to evaluate POCUS-hd for intrauterine pregnancy (IUP) detection compared to transabdominal and transvaginal ultrasound (TUTV), evaluate the inter-device agreement and inter-rater reliability of gestational age during early pregnancy.
Methods
It is an observational transverse study with consecutive patient recruitment. Two blinded operators systematically used POCUS-hd and reference transabdominal ultrasound for IUP diagnosis. The accuracy of POCUS-hd for IUP diagnosis was expressed as sensitivity, specificity, negative predictive value (NPV) and positive predictive value (PPV). The gestational age (GA) was assessed based on the crown-rump length. The reliability and agreement of gestational age evaluation were assessed by Bland-Altman plots, kappa statistic, and intraclass correlation coefficients (ICC).
Results
POCUS-hd compared to TU had a sensitivity of 95-100%, a specifcity of 90-100%, PPV of 95-100% and NPV of 90-100%. Inter-rater agreement for IUP detection using POCUS-hd was very good, kappa=1.0; CI95% [0.9-1.0]. The inter-device agreement limits (mean difference ± 2SD) for GA were: -3 to +2.3 days by Operator 1, -3.4 to +3.3 days by Operator 2 for POCUS-hd vs. TU and -3.1 to +2.3 days for POCUS-hd versus TUTV.
Conclusion
This handheld POCUS device is an accurate and reliable diagnostic tool that can be used for IUP positive findings and GA assessment during early pregnancy by clinicians in family planning settings or general practice.
Schlagwörter
Stichwörter der Autor:innen
Autor:innen und Institutionen
- Department of General Practice, Faculty of Medicine, Sorbonne University Paris France.
- Department of Adult Radiology, Necker Hospital Paris France.
Zitieren
APA
Skendi, M., Liard, R., Besacier, C., Correas, J. M., Moussaoui, S., Chastang, J., & Ibanez, G. (2022). Intrauterine Pregnancy Detection and Gestational Age Assessment During Early Pregnancy by a Handheld Point-Of-Care Ultrasound Device Compared to a High-End Ultrasound System. An Accuracy and Reliability Study. POCUS Journal, 7(2), 225-231. https://doi.org/10.24908/pocus.v7i2.15458
Vancouver
Skendi M, Liard R, Besacier C, Correas JM, Moussaoui S, Chastang J, et al. Intrauterine Pregnancy Detection and Gestational Age Assessment During Early Pregnancy by a Handheld Point-Of-Care Ultrasound Device Compared to a High-End Ultrasound System. An Accuracy and Reliability Study. POCUS Journal. 2022;7(2):225-231. doi:10.24908/pocus.v7i2.15458
BibTeX
@article{Skendi2022,
author = {Skendi, Mariela and Liard, Roxane and Besacier, Charlotte and Correas, Jean-Michel and Moussaoui, Sohela and Chastang, Julie and Ibanez, Gladys},
title = {Intrauterine Pregnancy Detection and Gestational Age Assessment During Early Pregnancy by a Handheld Point-Of-Care Ultrasound Device Compared to a High-End Ultrasound System. An Accuracy and Reliability Study},
journal = {POCUS Journal},
year = {2022},
volume = {7},
number = {2},
pages = {225--231},
doi = {10.24908/pocus.v7i2.15458}
}
EUVIMED