Assessing Diagnostic Accuracy of a Handheld Point-of-Care Device for Quantifying Neonatal Bilirubin Levels: The BEAT Jaundice @Home Study

Lauren E H Westenberg, Jasper V Been, Andrei Tintu, Jolande Y Vis, Helene A Bouma, Daan Nieboer, Peter H Dijk, Henk Groen, Marten J Poley, Erwin Ista, Eric A P Steegers, Irwin K M Reiss und 2 weitere

Neonatology · 11.03.2026 · 1-10

DOI: 10.1159/000551324 ↗

Kurz gefasst

Multizentrische Studie in der Zeitschrift Neonatology, online erschienen am 11.03.2026. Verfasst von Lauren E H Westenberg und 13 weiteren Autor:innen. Beteiligte Einrichtung: Division of Neonatology, Department of Neonatal and Pediatric Intensive Care, Erasmus MC Sophia Children's Hospital, University Medical Centre Rotter…. Hauptthemen laut MEDLINE: Bilirubin, Jaundice, Neonatal, Neonatal Screening, Point-of-Care Systems. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

Introduction

The Bilistick is a handheld point-of-care device for measuring total bilirubin levels in small blood volumes. We assessed its diagnostic accuracy and user convenience in near-term neonates cared for at home.

Methods

A prospective cohort study was conducted in nine Dutch community midwifery practices. Neonates ≥35 weeks’ gestation were eligible if they were at home between postnatal days 2–8 and had not received phototherapy. A Bilistick version 2.0 was used in parallel to laboratory-based bilirubin (LBB) quantification when significant visible jaundice was observed or the transcutaneous bilirubin reading was elevated.

Results

A total of 2,314 neonates were included in the study, with 423 blood samples analyzed across 13 laboratories. On 203 occasions, the Bilistick was not used. Among the remaining 220 Bilistick readings, 104 failed, and 2 lacked corresponding LBB results. A Bland-Altman plot of 114 paired measurements of Bilistick and LBB showed a mean difference of +9.7 µmol/L (0.57 mg/dL) with corresponding 95% limits of agreement of −179.7 to +199.2 µmol/L (−10.5 to 11.7 mg/dL). The positive predictive value of a Bilistick reading for having a total serum bilirubin level above the phototherapy threshold was 36.4%. The negative predictive value was 90.1%, sensitivity 60%, and specificity 77.6%. Hemolysis (24%) contributed to overestimations by the Bilistick. Community midwives expressed multiple barriers related to user convenience.

Conclusion

Diagnostic accuracy of the Bilistick when used in the home setting was limited. Its use was further hindered by a significant proportion of failed readings and low user convenience when operated by midwives.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

BilirubinJaundice, NeonatalNeonatal ScreeningPoint-of-Care SystemsFemaleHumansInfant, NewbornMaleNetherlandsPredictive Value of TestsProspective StudiesReproducibility of ResultsCohort Studies

Stichwörter der Autor:innen

Diagnostic accuracyNeonatal bilirubinPoint of care

Substanzen

Bilirubin

Autor:innen und Institutionen

  • Division of Neonatology, Department of Neonatal and Pediatric Intensive Care, Erasmus MC Sophia Children's Hospital, University Medical Centre Rotterdam, Rotterdam, The Netherlands, l.westenberg@erasmusmc.nl.
  • Department of Obstetrics and Gynecology, Erasmus MC Sophia Children's Hospital, University Medical Centre Rotterdam, Rotterdam, The Netherlands, l.westenberg@erasmusmc.nl.
  • Division of Neonatology, Department of Neonatal and Pediatric Intensive Care, Erasmus MC Sophia Children's Hospital, University Medical Centre Rotterdam, Rotterdam, The Netherlands.
  • Department of Obstetrics and Gynecology, Erasmus MC Sophia Children's Hospital, University Medical Centre Rotterdam, Rotterdam, The Netherlands.
  • Department of Clinical Chemistry, Erasmus MC, University Medical Centre Rotterdam, Rotterdam, The Netherlands.
  • Division of Neonatology, Department of Pediatrics, Beatrix Children's Hospital, University Medical Centre Groningen, University of Groningen, Groningen, The Netherlands.

Zitieren

APA

Westenberg, L. E. H., Been, J. V., Tintu, A., Vis, J. Y., Bouma, H. A., Nieboer, D., Dijk, P. H., Groen, H., Poley, M. J., Ista, E., Steegers, E. A. P., Reiss, I. K. M., van der Geest, B. A. M., & Hulzebos, C. V. (2026). Assessing Diagnostic Accuracy of a Handheld Point-of-Care Device for Quantifying Neonatal Bilirubin Levels: The BEAT Jaundice @Home Study. Neonatology, 1-10. https://doi.org/10.1159/000551324

Vancouver

Westenberg LEH, Been JV, Tintu A, Vis JY, Bouma HA, Nieboer D, et al. Assessing Diagnostic Accuracy of a Handheld Point-of-Care Device for Quantifying Neonatal Bilirubin Levels: The BEAT Jaundice @Home Study. Neonatology. 2026:1-10. doi:10.1159/000551324

BibTeX
@article{Westenberg2026,
  author = {Westenberg, Lauren E H and Been, Jasper V and Tintu, Andrei and Vis, Jolande Y and Bouma, Helene A and Nieboer, Daan and Dijk, Peter H and Groen, Henk and Poley, Marten J and Ista, Erwin and Steegers, Eric A P and Reiss, Irwin K M and van der Geest, Berthe A M and Hulzebos, Christian V},
  title = {Assessing Diagnostic Accuracy of a Handheld Point-of-Care Device for Quantifying Neonatal Bilirubin Levels: The BEAT Jaundice @Home Study},
  journal = {Neonatology},
  year = {2026},
  pages = {1--10},
  doi = {10.1159/000551324}
}
Kostenlos starten →