A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats

Kellie Cammarano, Patrick C Carney, Hailey E Rose, Michael Merkhassine, Elizabeth Ippolito, Anne Arsenault, Noah M Plotsker, John P Loftus iD, Marcin Kowalczyk, Jennifer M B Prieto iD

American Journal of Veterinary Research · 2024

DOI: 10.2460/ajvr.24.05.0146 ↗

Kurz gefasst

Fachartikel in der Zeitschrift American Journal of Veterinary Research, 2024. Verfasst von Kellie Cammarano und 9 weiteren Autor:innen. Beteiligte Einrichtung: Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY. Hauptthemen laut MEDLINE: Blood Glucose, Point-of-Care Systems. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 22 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY-NC 4.0).

AbstractSchlagwörterBibliografische AngabenZitierenVerwandte Studien

Abstract (Originalsprache)

OBJECTIVE To determine the clinical and analytical accuracy of a new veterinary-calibrated portable blood glucose monitor (PBGM) compared to a reference laboratory analyzer. ANIMALS Client-owned dogs (n = 77) and cats (n = 64). METHODS Peripheral and paired capillary whole-blood glucose concentrations measured via PBGM were compared to plasma glucose concentrations measured via a Cobas c501 reference analyzer (Roche). Analytical accuracy was evaluated with the Spearman rank correlation coefficient, Bland-Altman difference plot analysis, and Deming regression. Clinical accuracy was evaluated with Parkes error grid analysis. Paired peripheral and capillary blood samples were compared with the Wilcoxon matched-pairs signed-rank test. RESULTS There was a high correlation between PBGM and reference analyzer readings in dogs and cats. Human quality assurance standards (International Organization for Standardization 15197:2013 guidelines) for analytical accuracy were met for 95% of feline peripheral blood samples and 89% of canine samples. Similar veterinary standards (American Society of Veterinary Clinical Pathology guidelines) were met for 89% of canine and 92% of feline peripheral blood glucose measurements. Error grid analysis showed that all peripheral canine and 97% of feline measurements were clinically accurate (zone A). Any altered clinical decision for the remaining feline measurements was expected to minimally impact outcome (zone B). No significant difference was found between peripheral and capillary blood glucose measurements in either species. CLINICAL RELEVANCE The PBGM produced clinically accurate results and is suitable for use in veterinary and home settings to measure blood glucose.

Schlagwörter

MeSH-Schlagwörter (MEDLINE)

Blood GlucosePoint-of-Care SystemsDogsCatsAnimalsCat DiseasesDog DiseasesCalibrationFemaleMaleReproducibility of Results

Stichwörter der Autor:innen

catsclinical accuracydogsglucometerportable blood glucose monitor

Substanzen

Blood Glucose

Autor:innen und Institutionen

  • Department of Clinical Sciences, College of Veterinary Medicine, Cornell University, Ithaca, NY.
  • Universal Biosensors, Rowville, Victoria, Australia.

ORCID-Profile: John P Loftus, Jennifer M B Prieto

Zitieren

APA

Cammarano, K., Carney, P. C., Rose, H. E., Merkhassine, M., Ippolito, E., Arsenault, A., Plotsker, N. M., Loftus, J. P., Kowalczyk, M., & Prieto, J. M. B. (2024). A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats. American Journal of Veterinary Research. https://doi.org/10.2460/ajvr.24.05.0146

Vancouver

Cammarano K, Carney PC, Rose HE, Merkhassine M, Ippolito E, Arsenault A, et al. A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats. American Journal of Veterinary Research. 2024. doi:10.2460/ajvr.24.05.0146

BibTeX
@article{Cammarano2024,
  author = {Cammarano, Kellie and Carney, Patrick C and Rose, Hailey E and Merkhassine, Michael and Ippolito, Elizabeth and Arsenault, Anne and Plotsker, Noah M and Loftus, John P and Kowalczyk, Marcin and Prieto, Jennifer M B},
  title = {A veterinary-calibrated point-of-care glucometer accurately measures blood glucose concentration in dogs and cats},
  journal = {American Journal of Veterinary Research},
  year = {2024},
  doi = {10.2460/ajvr.24.05.0146}
}
Kostenlos starten →