A new tibial insert design with ball-in-socket medial conformity and posterior cruciate ligament retention has low tibial baseplate migration after unrestricted kinematically aligned total knee arthroplasty: a cohort study using radiostereometric analysis
Acta Orthopaedica · 2024
DOI: 10.2340/17453674.2024.42489 ↗Kurz gefasst
Kohortenstudie in der Zeitschrift Acta Orthopaedica, 2024. Verfasst von Abigail E Niesen und 3 weiteren Autor:innen. Beteiligte Einrichtung: epartment of Biomedical Engineering, University of California, Davis, CA, USA. Hauptthemen laut MEDLINE: Arthroplasty, Replacement, Knee, Radiostereometric Analysis, Prosthesis Design, Posterior Cruciate Ligament. Die Arbeit wurde bisher 11-mal zitiert (Crossref) und nennt 34 Literaturangaben. Der Volltext ist frei zugänglich (Open Access, CC BY 4.0).
Abstract (Originalsprache)
Background and purpose: In total knee arthroplasty (TKA), an insert with ball-in-socket (BS) medial conformity (MC) and posterior cruciate ligament (PCL) retention restores kinematics closer to native than an insert with intermediate (I) MC. However, high medial conformity might compromise baseplate stability as indicated by maximum total point motion (MTPM). Using the BS MC insert with PCL retention, we aimed to determine whether (i) the baseplate is stable as indicated by mean MTPM < 0.5 mm, (ii) baseplate stability is not strongly correlated to varus baseplate alignment, and (iii) baseplate stability, clinical outcome scores, and flexion are comparable with that of an I MC insert cohort which has demonstrated high stability, clinical outcome scores, and flexion.
Methods
Unrestricted kinematic alignment (unKA) TKA was performed on a cohort of 35 patients using a cemented baseplate. Biplanar radiographs acquired at timepoints up to 12 months were processed with model-based radiostereometric analysis (RSA) software to determine MTPM.
Results
At 1 year, mean MTPM of 0.35 mm was significantly below 0.5 mm (P < 0.001). MTPM was not strongly correlated to varus baseplate alignment up to 9° (r = 0.12, 95% confidence interval –0.22 to 0.44). Equivalence analyses revealed that MTPM, Forgotten Joint Score, Oxford Knee Score, and maximum flexion for the sBS MC insert were comparable with the I MC insert.
Conclusion
Using the new BS MC insert with PCL retention, the tibial baseplate was stable at the group level at 1 year. Baseplate stability was not strongly related to varus baseplate and limb alignment. Comparable patient-reported outcome scores and maximum flexion/extension at 1 year were shown between the 2 insert designs.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Autor:innen und Institutionen
- epartment of Biomedical Engineering, University of California, Davis, CA, USA.
- Department of Biomedical Engineering, University of California, Davis, CA, USA.
- Department of Biomedical Engineering, University of California, Davis, CA; Department of Mechanical Engineering, University of California, Davis, CA; Department of Orthopaedic Surgery, University of California Davis Med…
Zitieren
APA
Niesen, A. E., Tirumalai, P. A., Howell, S. M., & Hull, M. L. (2024). A new tibial insert design with ball-in-socket medial conformity and posterior cruciate ligament retention has low tibial baseplate migration after unrestricted kinematically aligned total knee arthroplasty: a cohort study using radiostereometric analysis. Acta Orthopaedica. https://doi.org/10.2340/17453674.2024.42489
Vancouver
Niesen AE, Tirumalai PA, Howell SM, Hull ML. A new tibial insert design with ball-in-socket medial conformity and posterior cruciate ligament retention has low tibial baseplate migration after unrestricted kinematically aligned total knee arthroplasty: a cohort study using radiostereometric analysis. Acta Orthopaedica. 2024. doi:10.2340/17453674.2024.42489
BibTeX
@article{Niesen2024,
author = {Niesen, Abigail E and Tirumalai, Pranav A and Howell, Stephen M and Hull, Maury L},
title = {A new tibial insert design with ball-in-socket medial conformity and posterior cruciate ligament retention has low tibial baseplate migration after unrestricted kinematically aligned total knee arthroplasty: a cohort study using radiostereometric analysis},
journal = {Acta Orthopaedica},
year = {2024},
doi = {10.2340/17453674.2024.42489}
}
EUVIMED