The corrosion products of proprietary and generic orthodontic fixed lingual retainers and their in-vitro cytotoxicity
The Angle Orthodontist · 2024
DOI: 10.2319/112823-788.1 ↗Kurz gefasst
Fachartikel in der Zeitschrift The Angle Orthodontist, 2024. Verfasst von Nessa A Finlay und 6 weiteren Autor:innen. Hauptthemen laut MEDLINE: Orthodontic Retainers, Fibroblasts, Materials Testing. Die Arbeit wurde bisher 1-mal zitiert (Crossref) und nennt 24 Literaturangaben. Der Volltext ist frei in PubMed Central verfügbar.
Abstract (Originalsprache)
Objective
To assess the corrosion products and cytotoxicity of generic and proprietary fixed lingual retainers (FLRs). Materials and Methods Seven FLRs were investigated. Wires were submersed in solution for 34 days, at 37°C, under constant agitation. A proportion of this solution was analyzed to determine the concentration of metallic ions leaching off the wires. The remainder was diluted to 5%, 10% and 20% followed by exposure to human gingival fibroblasts and analysis of cytotoxicity of the wires.
Results
Three wires (Dentaflex, Universal, and AZDent) released excessive concentrations of lead, two wires (MeshMark and Orthoflex) released excessive concentrations of nickel, and one wire (Universal) released excessive concentrations of molybdenum into solution. No statistically significant difference was found between the wires analyzed ( P = .24). Slight cytotoxicity was noted in only one wire (Dentaflex) at a 20% dilution of eluent. This was also the wire which released the highest concentration of lead into solution. All other wires, at all concentrations, were deemed noncytotoxic, but five samples overall were deemed statistically significant ( P < .0024). A statistically significant difference existed between wires ( P = .013) and concentrations analyzed ( P < .001).
Conclusions
Metals were released in differing quantities from all wires, with some elemental concentrations measuring more than that deemed acceptable in drinking water in Australia. A trend toward increased cell viability across samples was found with only one demonstrating cytotoxicity. There was no indication that generic FLRs were more or less biocompatible than their proprietary counterparts.
Schlagwörter
MeSH-Schlagwörter (MEDLINE)
Stichwörter der Autor:innen
Substanzen
Zitieren
APA
Finlay, N. A., Cheng, L., Kelly, E., Petocz, P., Gandedkar, N., Darendeliler, M. A., & Dalci, O. (2024). The corrosion products of proprietary and generic orthodontic fixed lingual retainers and their in-vitro cytotoxicity. The Angle Orthodontist. https://doi.org/10.2319/112823-788.1
Vancouver
Finlay NA, Cheng L, Kelly E, Petocz P, Gandedkar N, Darendeliler MA, et al. The corrosion products of proprietary and generic orthodontic fixed lingual retainers and their in-vitro cytotoxicity. The Angle Orthodontist. 2024. doi:10.2319/112823-788.1
BibTeX
@article{Finlay2024,
author = {Finlay, Nessa A and Cheng, Lam and Kelly, Elizabeth and Petocz, Peter and Gandedkar, Narayan and Darendeliler, Mehmet Ali and Dalci, Oyku},
title = {The corrosion products of proprietary and generic orthodontic fixed lingual retainers and their in-vitro cytotoxicity},
journal = {The Angle Orthodontist},
year = {2024},
doi = {10.2319/112823-788.1}
}
EUVIMED