Automated Lewis Score Estimation for Small Bowel Crohn Disease Using Capsule Endoscopy: Retrospective Cohort Study
JMIR Medical Informatics · 2026
DOI: 10.2196/92416 ↗At a glance
Kohortenstudie in JMIR Medical Informatics, 2026. Written by Yingying Meng and 3 co-authors. Institution: Children's Hospital of Fudan University, Shanghai, China. Major MEDLINE topics: Capsule Endoscopy, Crohn Disease, Intestine, Small. 22 references. The full text is freely available in PubMed Central.
Abstract (original language)
Background
Small bowel capsule endoscopy (SBCE) is widely used to evaluate small bowel involvement in Crohn disease (CD); however, manual Lewis score calculation is time-consuming and susceptible to interobserver variability.
Objective
This study aimed to develop and preliminarily validate a multitask deep learning system for automated Lewis score estimation in patients with small bowel CD undergoing SBCE.
Methods
We developed Eval-Model, a multitask framework integrating You Only Look Once (YOLO)–style lesion detection and localization for ulcers and edema, a stenosis classification branch, and a long short-term memory–based temporal regression module. The primary dataset comprised SBCE examinations from 814 patients at the Children’s Hospital of Fudan University. In this dataset, 725 (89.1%) cases were included for lesion-level annotation, and 89 (10.9%) cases with expert-consensus Lewis scores were used for scoring model development and internal testing. These 89 cases comprised 49 (55.1%) training cases and 40 (44.9%) internal test cases. All splits were performed at the patient level. A retrospective same-center temporal validation cohort of 20 consecutive cases from 2026 and 5 control cases were used for preliminary assessments of temporal robustness and specificity.
Results
In the 40-case internal Lewis score test set, Eval-Model achieved a mean squared error of 14.693 (95% CI 8.72-22.15), root mean squared error of 3.833 (95% CI 2.95-4.71), mean absolute error of 2.9 (95% CI 2.1-3.8), R2 of 0.82 (95% CI 0.70-0.90), and explained variance of 0.83 (95% CI 0.72-0.91). For lesion recognition, the model achieved an overall mean average precision of 0.94 with an inference time of 15 ms per frame. In the 20-case same-center temporal validation cohort, the mean squared error was 26.848 (95% CI 12.51-45.33), the root mean squared error was 5.182 (95% CI 3.54-6.73), the mean absolute error was 3.3 (95% CI 2.2-4.6), R2 rounded to 1.00 (95% CI 0.98-1.00), and the explained variance rounded to 1.00 (95% CI 0.98-1.00). These temporal estimates require cautious interpretation because of the small, same-center cohort.
Conclusions
Eval-Model demonstrated preliminary feasibility for automated Lewis score estimation from SBCE in small bowel CD. The system may serve as a standardized adjunct to expert review, but larger multicenter prospective validation with independent external annotation is required before clinical deployment.
Subjects
MeSH terms (MEDLINE)
Author keywords
Authors and institutions
- Children's Hospital of Fudan University, Shanghai, China.
- Chongqing Jinshan Technology (Group) Co, Ltd, Chongqing, China.
ORCID profiles: Yingying Meng, Min Gao, Jie Wu, Ying Huang
Cite
APA
Meng, Y., Gao, M., Wu, J., & Huang, Y. (2026). Automated Lewis Score Estimation for Small Bowel Crohn Disease Using Capsule Endoscopy: Retrospective Cohort Study. JMIR Medical Informatics. https://doi.org/10.2196/92416
Vancouver
Meng Y, Gao M, Wu J, Huang Y. Automated Lewis Score Estimation for Small Bowel Crohn Disease Using Capsule Endoscopy: Retrospective Cohort Study. JMIR Medical Informatics. 2026. doi:10.2196/92416
BibTeX
@article{Meng2026,
author = {Meng, Yingying and Gao, Min and Wu, Jie and Huang, Ying},
title = {Automated Lewis Score Estimation for Small Bowel Crohn Disease Using Capsule Endoscopy: Retrospective Cohort Study},
journal = {JMIR Medical Informatics},
year = {2026},
doi = {10.2196/92416}
}
EUVIMED